摘要

This paper describes modifications to an original design, correction factors and uncertainty evaluations for an improved free-air ionization chamber constructed at the Institute of Nuclear Energy Research (INER, Taiwan). In addition, a comparison of secondary standard air kerma calibration coefficients for 100-250 kV medium-energy X-rays was performed to verify the experimental accuracy and measurement consistency of the improved chamber. The comparison results showed a satisfactory agreement in the measurements which were within the combined expanded uncertainties (k = 2). (C) 2004 Elsevier Ltd. All rights reserved.
机译:本文介绍了对由核能研究所(台湾国家核能研究所)建造的改进型自由空气电离室的原始设计,校正因子和不确定性评估的修改。此外,对100-250 kV中能X射线的次级标准空气比释动能校准系数进行了比较,以验证改进腔室的实验准确性和测量一致性。比较结果表明,在扩展的不确定性(k = 2)内,测量结果令人满意。 (C)2004 Elsevier Ltd.保留所有权利。