Air Kerma Calibration Factors and kch Values for PTW Soft X-ray, NACP and Roos Ionization Chambers at Very Low X-ray Energies (0.035 mm - 1.0 mm Al HVL)

机译:非常低X射线能量的pTW软X射线,NaCp和Roos电离室的空气Kerma校准系数和kch值(0.035 mm-1.0 mm al HVL)

获取原文

摘要

Several national and international protocols have been established for the dosimetry of x-ray beams used in radiotherapy. For the very low energy x-rays (0.035mm-1.0 mm Al HVL) only two codes are available: the UK IPEMB Code of Practice and the German standard, DIN 6809 Part 4. The measurement of very low energy x-ray beams is normally performed with parallel plate ionization chambers calibrated at a standards laboratory and characterized by an air kerma calibration factor N{sub k}. According to the IPEMB Code of Practice the absorbed dose in the user's beam should be determined by taking measurements with the parallel plate chamber positioned such that its entrance window is at the surface of a full-scatter water equivalent phantom. The absorbed dose to water can then be determined using an equation which includes a factor, k{sub ch}, which accounts for the change in response of the ionization chamber between the calibration in air and measurement at the surface of the phantom. N{sub k} and k{sub ch} values for the PTW soft X-ray, NACP and Roos ionization chambers are reported. It was found that k{sub ch} values varied from about 1.01 to 1.08 depending on the chamber, beam quality and phantom material. It is recommended that the IPEMB Code of Practice should be revised to incorporate these values.

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号