首页> 外文期刊>Radiation measurements >The effect of holder design on the response of the HPA neutron personal dosemeter
【24h】

The effect of holder design on the response of the HPA neutron personal dosemeter

机译:支架设计对HPA中子个人剂量仪响应的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Modifications to the holder of the HPA neutron personal dosemeter are proposed which make the etched surface of the dosemeter have an effective depth equivalent to 10 mm of ICRU 4-element tissue. The effect of these changes has been investigated using Monte Carlo modelling with the code MCNP5. These calculations have been used to determine the energy dependence of the reading per unit H-p(10) for the current and modified designs, with the resultant response characteristics then being folded with workplace field energy distributions to determine the success or otherwise of the proposed changes.
机译:提出了对HPA中子个人剂量仪支架的修改,使剂量仪的蚀刻表面具有等效于ICRU 4元素组织的10毫米的有效深度。已经使用代码为MCNP5的Monte Carlo建模研究了这些更改的影响。这些计算已用于确定当前设计和修改后设计的每单位H-p(10)读数的能量依赖性,然后将结果响应特性与工作场所现场能量分布进行折叠,以确定拟议更改的成功与否。