Characteristics study of clear polymethylmethacrylate dosimeter, Radix W, in several kGy range

机译:透明聚甲基丙烯酸甲酯剂量计Radix W在数kGy范围内的特性研究

获取原文
获取原文并翻译 | 示例
       

摘要

Basic characteristics of Radix W, a commercially available undyed polymethylmethacrylate (PMMA) dosimeter conventionally used by readout at 320 nm, were studied in the dose range of 0.5-8 kGy, for its wide application especially for the evaluation of the sterilization dose and the quality assurance of food irradiation. The characteristics of dose response, the effect of irradiation temperature, and its stability after irradiation were examined over candidate readout wavelengths of 270-320 nm. The dose response readout at shorter wavelength is higher than that at longer wavelength, and 280 nm is the suitable readout wavelength for measurement of dose range of 0.5-8 kGy. The post-irradiation stability of dose response for 6 kGy is less than 1% within 24 h after irradiation at an irradiation temperature of 20 °C. Dose response is higher with temperature at irradiation temperatures in the range of -40 to 20 °C
机译:Radix W的基本特性是在0.5-8 kGy的剂量范围内进行研究的,它通常是在320 nm的读数范围内使用的市售未染色聚甲基丙烯酸甲酯(PMMA)剂量计,其广泛应用,尤其是用于评估灭菌剂量和质量确保食物辐照。在270-320 nm的候选读数波长上检查了剂量响应的特性,辐照温度的影响及其辐照后的稳定性。较短波长处的剂量响应读数高于较长波长处的剂量响应读数,并且280 nm是测量0.5-8 kGy剂量范围的合适读数波长。在20°C的照射温度下照射后24小时内,6 kGy剂量响应的照射后稳定性低于1%。辐射温度在-40至20°C范围内时,剂量响应会更高