A combined use of the BD-PND and BDT bubble detectors in neutron dosimetery

机译:BD-PND和BDT气泡探测器在中子剂量学中的结合使用

获取原文
获取原文并翻译 | 示例

摘要

The BDT bubble detector is predominantly sensitive to thermal neutrons and the BD-PND bubble detector is only sensitive to neutrons with energies larger than about 50 keV. The response of both types of bubble detectors was measured for four radionuclide sources, for thermal neutrons and for several mono-energetic neutron beams. At the same time the sensitivity given by the manufacturer was checked. Combining the two measured sensitivity functions with adequate weighing factors provides a better neutron dose equivalent response for the entire energy range. The effect of the new ICRP conversion factors (ICRP, 1996) following the ICRP 60 recommendations (ICRP, 1990) is determined. (C) 1998 Elsevier Science Ltd. All rights reserved. [References: 12]
机译:BDT气泡探测器主要对热中子敏感,而BD-PND气泡探测器仅对能量大于约50 keV的中子敏感。对四种放射性核素源,热中子和几个单能中子束,测量了两种类型的气泡探测器的响应。同时检查制造商给出的灵敏度。将两个测得的灵敏度函数与足够的加权因子结合在一起,可以在整个能量范围内提供更好的中子剂量当量响应。确定了遵循ICRP 60建议(ICRP,1990)的新ICRP转换因子(ICRP,1996)的效果。 (C)1998 Elsevier ScienceLtd。保留所有权利。 [参考:12]