首页> 外文期刊>Respiration: International Review of Thoracic Diseases >How to modify a volume-cycled home ventilator to satisfy a child's need for pressure-limiting and continuous positive airway pressure during spontaneous breathing.
【24h】

How to modify a volume-cycled home ventilator to satisfy a child's need for pressure-limiting and continuous positive airway pressure during spontaneous breathing.

机译:如何修改容积循环的家用呼吸机,以满足儿童在自发呼吸过程中对压力限制和持续正气道压力的需求。

获取原文
获取原文并翻译 | 示例
       

摘要

Piston driven volume-cycled home ventilators increase work of breathing in the synchronized intermittent mandatory ventilation mode. A 2-year-old trisomy 21 patient with chronic lung disease due to recurrent aspiration pneumonia required a ventilator rate of 6-8 breaths per minute awake and 15 asleep, with peak pressure of 32 cm H2O and positive end-expiratory pressure (PEEP) of 10 cm H2O. Two circuits were designed to facilitate breathing and respiratory mechanics of his spontaneous breaths on both were compared. A spring valve on the inspiratory line pressure-limited ventilation and at the end of the expiratory line provided PEEP in both systems. The reservoir system had a 2-liter bag on the inspiratory limb of the circuit, and two one-way valves at the patient connector to direct inspiratory and expiratory flow. The continuous positive airway pressure (CPAP) system provided continuous flow with a CPAP device set to deliver a pressure 2 cmH2O higher than the PEEP valve. On the CPAP system, compared to the reservoir system, dynamic compliance was greater [1.52 (0.14 SD) ml/cm H2O/kg vs. 0.39 (0.02), p < 0.001] and resistance less [8.15 (1.26) cm H2O/l/s vs. 45.86 (0.87), p < 0.001] as measured with a PeDS machine. This is an innovative use of a CPAP device.
机译:在同步间歇强制通风模式下,活塞驱动的容积循环式家用通风机增加了呼吸功。一名2岁,因反复吸入性肺炎而患有慢性肺病的21三体病患者需要通气速率为每分钟清醒时6-8次呼吸和15次入睡,峰值压力为32 cm H2O和呼气末正压(PEEP) 10厘米水柱设计了两个回路以促进呼吸,并比较了两个回路的自发呼吸的呼吸机制。在两个系统中,吸气管路限压通气和呼气管路末端的弹簧阀都提供了PEEP。储液器系统在回路的吸气肢体上有一个2升的袋子,在患者连接器处有两个单向阀,用于引导吸气和呼气流量。持续的气道正压(CPAP)系统通过CPAP装置提供连续的流量,该装置设置为输送的压力比PEEP阀高2 cmH2O。在CPAP系统上,与水库系统相比,动态顺应性更高[1.52(0.14 SD)ml / cm H2O / kg,而0.39(0.02),p <0.001],阻力较小[8.15(1.26)cm H2O / l / s对45.86(0.87),p <0.001](用PeDS机器测量)。这是CPAP设备的创新用法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号