The use of a multipurpose phantom for mailed dosimetry checks of therapeutic photon beams: 'OPERA' (operational phantom for external radiotherapy audit).

机译:使用多功能体模邮寄治疗性光子束的剂量学检查:“ OPERA”(用于外部放射疗法审核的操作体模)。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND PURPOSE: This study presents a technical description of the solid multipurpose phantom (MPP) 'OPERA' (operational phantom for external radiotherapy audit) for mailed dosimetry checks of photon beams in reference and non-reference conditions. The aim is to test the functionality of this phantom by comparing the results provided by the local treatment planning system (TPS) with the measurements by film and thermoluminescence dosimeters (TLDs) in a number of clinical conditions.MATERIAL AND METHODS: The polystyrene MPP is constructed to check the following five irradiation conditions: square fields, asymmetrical fields, wedged beams, oblique incidence and influence of inhomogeneities in the field. The absorbed dose on the central beam axis is measured with TLDs for the first three irradiation conditions and the relative dose distributions are verified with film.RESULTS: The 'OPERA' phantom and the corresponding instruction sheets were mailed to radiotherapy centres of 12 different countries, in order to verify its functionality. A good agreement between TPS and the film data is found: the mean deviations between relative dose profiles obtained from film and locally applied TPS data range between -1.2 and 0.4%, except for the short side of the asymmetrical field. The standard deviations (SDs) generally increase off-axis, with a large SD on the asymmetrical side. Considering the TLD results, a mean ratio of measured to stated dose of 0.995+/-0.014 (1 SD), 0.988+/-0.019 (1 SD) and 0.981+/-0.024 (1 SD) for set-up 1 (square field), 2 (asymmetrical field) and 3 (wedged field), respectively, has been found.CONCLUSION: The 'OPERA' phantom can be useful for on- and off-axis verification of the TPS, as well as for mailed dosimetrical checks of photon beams in reference and non-reference conditions.
机译:背景与目的:本研究介绍了在参考和非参考条件下,用于邮寄剂量学检查的实心多用途体模(MPP)'OPERA'(用于外部放射疗法审核的操作体模)的技术说明。目的是通过比较局部治疗计划系统(TPS)提供的结果与胶片和热致发光剂量计(TLD)在许多临床条件下的测量结果来测试该体模的功能。材料与方法:聚苯乙烯MPP为构造为检查以下五个照射条件:方形场,非对称场,楔形光束,斜入射和场中不均匀性的影响。在前三个辐照条件下,使用TLD测​​量中心光束轴上的吸收剂量,并用胶片验证相对剂量分布。结果:“ OPERA”体模和相应的说明书已邮寄到12个不同国家的放射治疗中心,为了验证其功能。在TPS和胶片数据之间找到了很好的一致性:从胶片获得的相对剂量曲线与局部施加的TPS数据之间的平均偏差在-1.2%和0.4%之间,除了不对称场的短边。通常,标准偏差(SD)离轴增加,而非对称侧的SD较大。考虑到TLD结果,设置1(平方)的实测剂量与规定剂量的平均比率为0.995 +/- 0.014(1 SD),0.988 +/- 0.019(1 SD)和0.981 +/- 0.024(1 SD)结论:“ OPERA”模型对于TPS的轴上和轴外验证以及邮寄的剂量学检查非常有用。参考和非参考条件下光子束的分布

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号