Display systems are key components of the digital radiology department. Current display systems for medical imaging are based on cathode-ray tubes (CRTs) or active-matrix liquid crystal displays (AMLCDs). The CRT is a cathodoluminescent display: Light is generated by exciting a luminescent material with energetic electrons. AMLCDs are light-modulating devices that form the image in the screen by controlling the transparency of individual display pixels. Many image quality aspects of CRTs are determined by the way the pixel luminance is generated in the cathodoluminescent screen. The resolution properties of AMLCDs are much better than those of CRTs. In CRT devices, phosphor granularity and raster scanning patterns are the main components of spatial noise. In AMLCDs, the most notable feature of the noise characteristic is the subpixel structure of complex pixel designs used in medical displays. The small-spot contrast of CRTs is dominated mainly by veiling glare and reflections of ambient illumination. In addition to display reflectance, the contrast of medical AMLCDs is affected by crosstalk and by variations of the luminance at off-normal viewing angles.
机译:显示系统是数字放射科的关键组件。当前用于医学成像的显示系统基于阴极射线管(CRT)或有源矩阵液晶显示器(AMLCD)。 CRT是阴极发光显示器:通过用高能电子激发发光材料来产生光。 AMLCD是通过控制各个显示像素的透明度在屏幕上形成图像的光调制设备。 CRT的许多图像质量方面取决于在阴极荧光屏中生成像素亮度的方式。 AMLCD的分辨率特性比CRT更好。在CRT设备中,磷光体粒度和光栅扫描图案是空间噪声的主要成分。在AMLCD中,噪声特性最显着的特征是医疗显示器中使用的复杂像素设计的子像素结构。 CRT的小点对比度主要由遮盖的眩光和环境照明的反射决定。除显示反射率外,医疗AMLCD的对比度还受串扰和偏离正常视角的亮度变化的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号