Factors determining the performance of triple quadrupole, quadrupole ion trap and sector field mass spectrometers in electrospray ionization tandem mass spectrometry of oligonucleotides. 1. Comparison of performance characteristics

机译:决定三重四极杆,四极杆离子阱和扇形场质谱仪在寡核苷酸电喷雾电离串联质谱分析中的性能的因素。 1.性能特点比较

获取原文
获取原文并翻译 | 示例
       

摘要

The performance of triple-stage quadrupole (TSQ), quadrupole ion trap (QIT), and double focusing sector field (DFSF) mass spectrometers for the generation of fragment ions to obtain sequence information about oligonucleotides was compared. Upon electrospray ionization (ESI), the charge-state distribution of candidate precursor ions not only varied significantly with the type of mass spectrometer, but also with the size and sequence of the investigated oligonucleotides. While concentration limits of detection for an octanucleotide were in the 100 pmol/L range on the QIT and in the 5-10 nmol/L range on the TSQ and DFSF instruments, those of a 24-mer were in the 2-13 nmol/L range on all three instruments. Reproducibility of mass determination, an important prerequisite for reliable identification of fragment ions, was highest on the TSQ with 0.0037% relative standard deviation over three days. Finally, the tandem mass spectra of a dimethoxytritylated pentanucleotide recorded on the three instruments were compared. Relatively simple spectra dominated by complete series of fragment ions of the (a-B) and w type were obtained on the QIT. Complete series of (a-B) and w ions were also observed on the TSQ. However, additional fragments belonging to the b,c,d,x and z series were found in the spectrum. In the spectrum recorded after in-source fragmentation in the DFSF, only fragments corresponding to the loss of a nucleobase and a complete series of w ions were observed. All three mass spectrometers were suitable for the generation of fragment ions, from which the complete nucleotide sequence of the pentanucleotide could be deduced. Copyright (C) 2001 John Wiley & Sons, Ltd. [References: 31]
机译:比较了三级四极杆(TSQ),四极离子阱(QIT)和双聚焦扇形场(DFSF)质谱仪产生碎片离子以获得寡核苷酸序列信息的性能。在电喷雾电离(ESI)时,候选前体离子的电荷状态分布不仅随质谱仪的类型而变化,而且随所研究的寡核苷酸的大小和序列而变化。虽然QIT的八核苷酸检测浓度限在100 pmol / L范围内,而TSQ和DFSF仪器的检测限在5-10 nmol / L范围内,但24聚体的检测限在2-13 nmol / L之间。所有三种乐器的L量程。质量测定的可重复性是可靠鉴定碎片离子的重要先决条件,在TSQ上三天的相对标准偏差为0.0037%,最高。最后,比较了三种仪器上记录的二甲氧基三苯甲基化的五核苷酸的串联质谱。在QIT上获得了由(a-B)和w型碎片离子的完整序列主导的相对简单的光谱。在TSQ上也观察到了完整的(a-B)和w离子系列。但是,在光谱中发现了属于b,c,d,x和z系列的其他片段。在DFSF中进行源内裂解后记录的光谱中,仅观察到与核碱基丢失和完整的w离子序列相对应的碎片。所有这三种质谱仪均适用于碎片离子的产生,从中可以推断出五核苷酸的完整核苷酸序列。版权所有(C)2001 John Wiley&Sons,Ltd. [引用:31]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号