Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography.

机译:用于数字胸部放射线照相的非晶硅平板探测器的成像特性。

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: To evaluate the imaging characteristics of an amorphous silicon flat-panel detector (FPD) for digital chest radiography. MATERIALS AND METHODS: The 41 x 41-cm digital FPD is constructed on a single monolithic glass substrate with a structured cesium iodide scintillator layer and an amorphous silicon thin-film transistor array for image readout. Basic imaging characteristics of the FPD and associated image processing system were assessed on acquired images, including linearity, repeatability, uniformity of response, modulation transfer function (MTF), noise power spectrum, detective quantum efficiency (DQE), contrast sensitivity, and scatter content. Results with the FPD system were compared to those with a storage phosphor computed radiography (CR) system. RESULTS: Images obtained with the FPD demonstrated excellent uniformity, repeatability, and linearity, as well as MTF and DQE that were superior to those with the storage phosphor CR system. The contrast and scatter content of images acquired with the FPD were equivalent to those acquired with the storage phosphor system. CONCLUSION: The FPD provides radiographic images with excellent inherent physical image quality.
机译:目的:评估用于数字胸部放射线照相的非晶硅平板探测器(FPD)的成像特性。材料与方法:41 x 41厘米数字FPD构建在具有结构化碘化铯闪烁体层和用于图像读取的非晶硅薄膜晶体管阵列的单个整体玻璃基板上。在获取的图像上评估了FPD和相关图像处理系统的基本成像特性,包括线性,可重复性,响应均匀性,调制传递函数(MTF),噪声功率谱,检测量子效率(DQE),对比度灵敏度和散射含量。将FPD系统的结果与存储磷光X线摄影(CR)系统的结果进行比较。结果:FPD获得的图像显示出优异的均匀性,可重复性和线性度,以及MTF和DQE优于存储荧光粉CR系统。使用FPD采集的图像的对比度和散射含量与使用荧光粉系统采集的图像的对比度和散射含量相同。结论:FPD为放射照相图像提供了出色的固有物理图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号