首页> 外文期刊>Radiology >Digital breast tomosynthesis: Observer performance of clustered microcalcification detection on breast phantom images acquired with an experimental system using variable scan angles, angular increments, and number of projection views
【24h】

Digital breast tomosynthesis: Observer performance of clustered microcalcification detection on breast phantom images acquired with an experimental system using variable scan angles, angular increments, and number of projection views

机译:数字化乳房断层合成:使用可变扫描角度,角度增量和投影视图数量的实验系统采集的乳房幻影图像上的集群微钙化检测的观察器性能

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To investigate the dependence of microcalcification cluster detectability on tomographic scan angle, angular increment, and number of projection views acquired at digital breast tomosynthesis (DBT).Materials and Methods: A prototype DBT system operated in step-and-shoot mode was used to image breast phantoms. Four 5-cm-thick phantoms embedded with 81 simulated microcalcification clusters of three speck sizes (subtle, medium, and obvious) were imaged by using a rhodium target and rhodium filter with 29 kV, 50 mAs, and seven acquisition protocols. Fixed angular increments were used in four protocols (denoted as scan angle, angular increment, and number of projection views, respectively: 16°, 1°, and 17; 24°, 3°, and nine; 30°, 3°, and 11; and 60°, 3°, and 21), and variable increments were used in three (40°, variable, and 13; 40°, variable, and 15; and 60°, variable, and 21). The reconstructed DBT images were interpreted by six radiologists who located the microcalcification clusters and rated their conspicuity.Results: The mean sensitivity for detection of subtle clusters ranged from 80% (22.5 of 28) to 96% (26.8 of 28) for the seven DBT protocols; the highest sensitivity was achieved with the 16°, 1°, and 17 protocol (96%), but the difference was significant only for the 60°, 3°, and 21 protocol (80%, P < .002) and did not reach significance for the other five protocols (P = .01-.15). The mean sensitivity for detection of medium and obvious clusters ranged from 97% (28.2 of 29) to 100% (24 of 24), but the differences fell short of significance (P = .08 to ..99). The conspicuity of subtle and medium clusters with the 16°, 1°, and 17 protocol was rated higher than those with other protocols; the differences were significant for subtle clusters with the 24°, 3°, and nine protocol and for medium clusters with 24°, 3°, and nine; 30°, 3°, and 11; 60°, 3° and 21; and 60°, variable, and 21 protocols (P < .002)Conclusion: With imaging that did not include x-ray source motion or patient motion during acquisition of the projection views, narrow-angle DBT provided higher sensitivity and conspicuity than wide-angle DBT for subtle microcalcification clusters.
机译:目的:研究微钙化团簇的可检测性对断层扫描角度,角度增量和在数字乳腺断层合成(DBT)上获得的投影视图数量的依赖性。材料与方法:使用了一种分步拍摄模式的DBT原型系统拍摄乳房幻像。使用29 kV,50 mAs和7个采集方案的铑靶和铑滤光片,对嵌入了81个模拟的微钙化簇的4个5厘米厚的幻像进行成像,这些簇分别具有三种斑点大小(细微,中等和明显)。在四种方案中使用了固定的角度增量(分别表示为扫描角,角度增量和投影视图数:16°,1°和17; 24°,3°和9; 30°,3°和11;和60°,3°和21),并且以三个增量(40°,可变和13; 40°,可变和15;以及60°,可变和21)使用可变增量。六位放射科医生对重建的DBT图像进行了解释,他们确定了微钙化簇并对其显着性进行了评估。结果:对于七个DBT,检测细微簇的平均灵敏度范围为80%(28。的22.5)到96%(28。的26.8)。协议;在16°,1°和17度实验方案中可获得最高灵敏度(96%),但差异仅在60°,3°和21度实验方案中显着(80%,P <.002)达到其他五种协议的重要性(P = .01-.15)。中度和明显簇的平均检测灵敏度在97%(29的28.2)到100%(24的24)之间,但是差异没有显着性(P = .08至..99)。 16°,1°和17协议的微妙和中等集群的显着性等级高于其他协议;对于使用24°,3°和9个协议的细微簇和使用24°,3°和9个协议的中型簇,差异是显着的; 30°,3°和11; 60°,3°和21;以及60°,可变和21种协议(P <.002)。结论:由于在获取投影视图期间不包括X射线源运动或患者运动的成像,与广角镜相比,窄角DBT具有更高的灵敏度和显眼性。微细钙化簇的角度DBT。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号