首页> 外文期刊>Radiology >Breast MR imaging at 3 t with dual-source radiofrequency transmission offers superior b1 homogeneity: An intraindividual comparison with breast mr imaging at 1.5 T1
【24h】

Breast MR imaging at 3 t with dual-source radiofrequency transmission offers superior b1 homogeneity: An intraindividual comparison with breast mr imaging at 1.5 T1

机译:双源射频传输的3 t乳腺MR成像具有出色的b1均匀性:与1.5 T1乳腺mr成像的个体差异

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose: To evaluate and compare B1 homogeneity for breast magnetic resonance (MR) imaging performed at 3 T with dualsource radiofrequency (RF) transmission to 1.5-T MR imaging and 3-T MR imaging with quadrature transmission. Materials and Methods: This prospective study received institutional review board approval and patients provided informed consent. Women (n = 25; mean age, 53 years; range, 30-68 years) suspected of having breast lesions underwent breast MR imaging examinations on comparable 1.5-T and 3-T clinical systems between February and May 2012. B1 maps were obtained at 1.5 T and at 3 T with quadrature and dualsource RF transmission. Intrabreast differences and differences in mean B1 values between right and left breasts were investigated by using two-sided multivariate analysis of variance with interaction; t tests were used to compare the differences between measured whole-breast mean B1 values and requested B1 values. Results: With quadrature transmission at 1.5 T and 3 T, the mean B1 values showed a statistically significant difference: leftbreast measured B1 was 28.9% of requested B1 value at 1.5 T and 213.7% at 3 T (P < .001), whereas right-breast measured B1 was +5.4% of requested B1 value at 1.5 T (P < .001) and +2.7% at 3 T (P = .01). With dual-source RF transmission at 3 T, mean B1 values across the breasts were not statistically different, nor were the measured B1 values compared with requested B1 values (left breast, 20.6%; right breast, 20.7%). At 3 T with dual-source transmission, slight intrabreast local variations in B1 were recorded. Conclusion: MR imaging at 3 T with dual-source RF transmission offered an overall B1 homogeneity for breast imaging that was better than that obtained at 1.5 T and with quadrature transmission.
机译:目的:评价和比较B1同质性在3 T双源射频(RF)传输与1.5-T MR成像和3-T MR正交传输成像中进行的胸部磁共振(MR)成像。材料和方法:该前瞻性研究获得了机构审查委员会的批准,患者获得了知情同意。在2012年2月至2012年5月之间,通过可比的1.5-T和3-T临床系统,对怀疑患有乳腺病变的女性(n = 25;平均年龄为53岁;范围为30-68岁)进行了MR成像检查。在1.5 T和3 T下具有正交和双源RF传输。通过使用交互作用方差的双向多变量分析,研究了乳房内差异和左右乳房之间的平均B1值差异。 t检验用于比较测得的整体乳房平均B1值和要求的B1值之间的差异。结果:在1.5 T和3 T正交传输时,平均B1值显示出统计学上的显着差异:左胸测得的B1是1.5 T时要求的B1值的28.9%,在3 T时为213.7%(P <.001),而右-乳房测得的B1在1.5 T时为要求的B1值的+ 5.4%(P <.001),在3 T时为+ 2.7%(P = .01)。在3 T的双源RF传输下,整个乳房的平均B1值在统计学上无差异,与所要求的B1值相比,测得的B1值也无统计学差异(左乳房为20.6%;右乳房为20.7%)。在双源传输的3 T处,记录到B1的乳房内局部微小变化。结论:双源RF传输在3 T下进行MR成像可提供总体的B1均匀性,比1.5 T和正交传输下获得的乳房成像更好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号