Effect of phase-encoding scheme on the relationship between contrast agent concentration and signal intensity on inversion recovery turbo fast low-angle shot T1-weighted images

机译:相位编码方案对反转恢复涡轮快速低角度拍摄T1加权图像上造影剂浓度和信号强度之间关系的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Previous studies have shown that the image parameters such as the repetition time and inversion time (TI) and image sequences can have an effect on the maximum linearity between signal intensity (SI) and concentration. Our aim in this study was to investigate the effect of a phase-encoding scheme (linear phase-encoding and center out phase-encoding) at different TIs on the maximum linear relationship between contrast agent concentration and SI using inversion recovery TurboFLASH (turbo fast low angle shot) T1-weighted images in MRI. A phantom was designed to hold 25 vials which contained either different (between 0 and 19.77 mmol/L) or constant (1.20 mmol/L) concentrations of contrast agent. The vials of constant concentration were used for measurement of coil non-uniformity. The vials of different concentrations were used for the measurement of the maximum linear relationship between SI and concentration and the effect of a phase-encoding scheme on the linearity, where the squared correlations (R2) between the corrected SI and the concentration were equal to 0.95 and 0.99. The results showed that the linear relationship between SI and concentration was dependent on the phase-encoding scheme and on the TIs. The results also showed that the maximum linearity with the linear phase-encoding acquisition was higher than that with the center out phase-encoding acquisition at the same R2. This study shows that the phase-encoding scheme is an important parameter when the SI is measured. These schemes can have an effect on the maximum linearity between SI and concentration.
机译:先前的研究表明,图像参数(例如重复时间和反转时间(TI))以及图像序列可能会影响信号强度(SI)和浓度之间的最大线性。我们在这项研究中的目的是使用反转恢复TurboFLASH(turbo fast low),研究不同TI上的相位编码方案(线性相位编码和居中相位编码)对造影剂浓度和SI之间最大线性关系的影响。角度拍摄)MRI中的T1加权图像。幻像被设计为容纳25个包含不同浓度(0到19.77 mmol / L)或恒定浓度(1.20 mmol / L)的造影剂的小瓶。用恒定浓度的小瓶测量线圈的不均匀性。不同浓度的小瓶用于测量SI与浓度之间的最大线性关系以及相位编码方案对线性的影响,其中校正后的SI与浓度之间的平方相关性(R2)等于0.95和0.99。结果表明,SI和浓度之间的线性关系取决于相位编码方案和TI。结果还表明,在相同的R2下,线性相位编码采集的最大线性度高于中心相位采集编码的最大线性度。这项研究表明,在测量SI时,相位编码方案是一个重要参数。这些方案可能会影响SI和浓度之间的最大线性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号