Effects of inversion and saturation times on relationships between contrast agent concentrations and signal intensities of T1-weighted magnetic resonance images.

机译:反转和饱和时间对T1加权磁共振图像的造影剂浓度和信号强度之间关系的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

The present study was an attempt to investigate the effect of variation of inversion time (T (I)) and saturation time (T (S)) on the linear relationship between contrast agent concentration and signal intensity (SI) on Turbo Fast Low Angle Shot (TurboFLASH) T (1)-weighted images in MRI. For this purpose, inversion recovery (IR) and saturation recovery (SR) sequences (Center out Phase-Encoding acquisition) were used. A phantom was designed to hold 25 vials which contained either different (between 0 and 19.77 mmol/L) or constant (1.20 mmol/L) concentrations of contrast agent. The vials of constant concentration were used for the measurement of coil non-uniformity, which was normalized to give a correction factor. The vials of different concentrations were used to measure the SI by using different sequences and different T (I) and T (S) values. To calculate the corrected SI for different concentrations, we multiplied the SI of each vial by its correction factor. The relationships between the corrected SI and the concentration [were evaluated], where the threshold of (R (2) = 0.95 and 0.99) was maintained. This study shows that different sequences and different T (I) and T (S) values can have an effect on the correlation between the SI and concentration. Regardless of the values of T (I), T (S), and the different IR and SR sequences chosen, the linear relationship between the SI and concentration was about twice that previously reported (i.e., 0.8 mmol/L, R (2) = 0.95).
机译:本研究试图研究反转时间(T(I))和饱和时间(T(S))的变化对Turbo快速低角度拍摄中造影剂浓度和信号强度(SI)线性关系的影响(TurboFLASH)MRI中的T(1)加权图像。为此,使用了反转恢复(IR)和饱和恢复(SR)序列(中心出相位编码采集)。幻像被设计为容纳25个包含不同浓度(0到19.77 mmol / L)或恒定浓度(1.20 mmol / L)的造影剂的小瓶。将恒定浓度的小瓶用于线圈不均匀性的测量,将其标准化以给出校正因子。通过使用不同的序列和不同的T(I)和T(S)值,使用不同浓度的小瓶来测量SI。为了计算不同浓度的校正后的SI,我们将每个小瓶的SI乘以校正因子。 [校正的SI]与浓度之间的关系[被评估],其中保持(R(2)= 0.95和0.99)的阈值。这项研究表明,不同的序列和不同的T(I)和T(S)值可能会影响SI与浓度之间的相关性。不管T(I),T(S)的值以及选择的不同IR和SR序列如何,SI与浓度之间的线性关系约为先前报道的两倍(即0.8 mmol / L,R(2) = 0.95)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号