Inter-Changeability of Impedance Devices for Lymphedema Assessment

机译:阻抗设备在淋巴水肿评估中的互换性

获取原文
获取原文并翻译 | 示例

摘要

Impedance technology is a popular technique for the early detection of lymphedema. The preferred approach is to use bioimpedance spectroscopy (BIS), with measurements being made with the subject lying supine, although attempts have been made to use single or multiple frequency impedance measurements obtained while the subject is standing. The aim of the present study was to determine the equivalence of these different approaches. Impedance measurements of the individual limbs of 37 healthy individuals were determined using both a stand-on, multi-frequency impedance device and a supine impedance spectroscopy instrument. Significant differences were found between the instruments in both absolute impedance values and, importantly, inter-limb impedance ratios. Since impedance ratios in healthy individuals provide the reference standard for detection of lymphedema, these data indicate that the methods are not interchangeable. Consideration of the errors associated with each method indicates that the BIS remains the preferred method for lymphedema detection.
机译:阻抗技术是一种用于早期检测淋巴水肿的流行技术。首选方法是使用生物阻抗谱(BIS),使受试者仰卧进行测量,尽管尝试使用站立时获得的单频或多频阻抗测量。本研究的目的是确定这些不同方法的等效性。使用立式多频阻抗设备和仰卧阻抗谱仪确定了37位健康个体的四肢的阻抗测量值。仪器之间在绝对阻抗值和重要的是肢间阻抗比方面都存在显着差异。由于健康个体的阻抗比为淋巴水肿的检测提供了参考标准,因此这些数据表明这些方法不可互换。考虑到与每种方法相关的错误,表明BIS仍然是淋巴水肿检测的首选方法。