The transthoracic impedance (T) and its variations may be estimated through the measurement of the electrical impedance between the can and the right ventricular coil of a defibrillation lead. This method may allow the monitoring of fluid overload before a heart failure attack. Aim of this study was to validate in vitro a method to calculate T in case of a standard bipolar pacing lead, by performing 3 measurements: standard unipolar impedance from the tip (Zuni-tip); unipolar impedance from the ring (Zuni-ring); standard bipolar impedance (Zbip). The formula we used is derived from the standard equivalent circuit of a pacing system: T = (Zuni-tip-Zbip + Zuni-ring)/2 T represents the tissue impedance between the can and the electrodes of the lead. To validate the method we used a saline solution and 3 different pacing leads manufactured by Vitatron (Vitatron BV, Arnhem, The Netherlands): Impulse II (high impedance lead), Crystalline ActFix (screw-in lead), Brilliant S+ (VDD single-lead). The measured values of the saline solution impedance were compared to the values calculated through the formula. Results: The calculated impedance of the solution, evaluated through the proposed formula, is reliable independently of the electrode used and highly correlated to the corresponding measured values (R > 0.9). Conclusion: Tissue impedance may be calculated from standard unipolar and bipolar impedance measurements with a standard bipolar pacing lead. Springer Science+Business Media, Inc. 2006.
机译:经胸阻抗(T)及其变化可通过测量除颤导线的罐与右心室线圈之间的电阻抗来估算。该方法可以允许在心力衰竭发作之前监测液体超负荷。这项研究的目的是通过进行3次测量,在体外验证在标准双极起搏导线情况下计算T的方法:尖端的标准单极阻抗(Zuni-tip);来自环的单极阻抗(Zuni-ring);标准双极阻抗(Zbip)。我们使用的公式源自起搏系统的标准等效电路:T =(Zuni-tip-Zbip + Zuni-ring)/ 2 T表示罐头和电极之间的组织阻抗。为了验证该方法,我们使用了盐溶液和Vitatron(Vitatron BV,阿纳姆,荷兰)制造的3种不同起搏引线:Impulse II(高阻抗引线),Crystalline ActFix(拧入引线),Brilliant S +(VDD单铅)。将盐溶液阻抗的测量值与通过公式计算出的值进行比较。结果:通过建议的公式评估的溶液阻抗计算结果是可靠的,与所使用的电极无关,并且与相应的测量值高度相关(R> 0.9)。结论:组织阻抗可通过标准双极起搏导线的标准单极和双极阻抗测量值计算得出。 Springer Science + Business Media,Inc. 2006。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号