The University of California (UC) Davis Reading Center evaluated 19,961 scans from 981 subjects in two multiple sclerosis therapeutic trials with the aim of determining the influence of optical coherence tomography quality control procedures on error rates. There was no optical coherence tomography technician certification in Trial 1, and technicians had very limited monitoring and feedback during the trial in view of the fact that data were received retrospectively. However, technicians were certified in Trial 2 and submitted data in accordance with the protocol. Trial 2 scans had higher signal strengths, fewer errors, and more useable data than the scans in Trial 1. Thus, certified technicians and prompt transmission of data for ongoing quality control monitoring provided higher quality data in multiple sclerosis trials.
机译:加利福尼亚大学戴维斯分校阅读中心在两项多发性硬化症治疗试验中对981名受试者的19,961项扫描进行了评估,目的是确定光学相干断层扫描质量控制程序对错误率的影响。在试验1中,没有光学相干断层扫描技术员资格证书,并且由于回顾性地接收到数据,技术员在试验期间的监控和反馈非常有限。但是,技术人员已在试验2中获得认证,并根据协议提交了数据。与第1次扫描相比,第2次扫描具有更高的信号强度,更少的错误和更多可用的数据。因此,经过认证的技术人员和即时数据传输以进行持续的质量控制监控,可在多次硬化试验中提供更高质量的数据。