首页> 外文期刊>Monatshefte fur Chemie >An optode membrane for determination of gold using a simple light-emitting diode-based device
【24h】

An optode membrane for determination of gold using a simple light-emitting diode-based device

机译:一种使用基于发光二极管的简单装置测定金的光电膜

获取原文
获取原文并翻译 | 示例
           

摘要

A selective optode based on immobilization of 5-(p-dimethylaminobenzylidene)rhodanine on a triacetyl-cellulose membrane was developed for quantitative determination of Au(III). The determination procedure was performed using a simple light-emitting diode (LED)-based device as a new effort to overcome low reproducibility and repeatability problems which usually accompany optode-based determinations. The results obtained were compared with those of conventional spectrophotometric methods. The response characteristics of the sensor including dynamic range, reproducibility, response time, and lifetime are discussed in detail. This sensor was used for the determination of Au(III) in ore and electroplating liquid effluent samples with satisfactory results in comparison with flame atomic absorption spectroscopy as standard method. Under the optimum conditions, the calibration plot was linear in the concentration range of 0.3-6.0 μg cm~(-3). The relative standard deviations for five replicate determinations of 1.0 μg cm~(-3) Au(III) and the corresponding limits of detection were 1.76% and 0.12 μg cm~(-3), respectively.
机译:开发了一种基于将5-(对二甲氨基苄叉基)罗丹宁固定在三乙酰纤维素膜上的选择性光电二极管,用于定量测定Au(III)。确定过程是使用基于简单发光二极管(LED)的设备执行的,作为克服通常伴随基于光电二极管的确定的低重现性和可重复性问题的一项新工作。将获得的结果与常规分光光度法进行比较。详细讨论了传感器的响应特性,包括动态范围,可重复性,响应时间和寿命。与标准方法火焰原子吸收光谱法相比,该传感器用于测定矿石和电镀废水中的Au(III),具有令人满意的结果。在最佳条件下,浓度范围为0.3-6.0μgcm〜(-3)时,校正曲线呈线性关系。五个重复测定的1.0μgcm〜(-3)Au(III)的相对标准偏差和相应的检出限分别为1.76%和0.12μgcm〜(-3)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号