Detection of gastric slow wave uncoupling from multi-channel electrogastrogram: validations and applications.

机译:从多通道胃电图检查胃慢波解耦:验证和应用。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Current methodology of single channel electrogastrography is unable to detect coupling or uncoupling of gastric slow waves, which is crucial for gastric emptying. In this study, a new methodology, called cross-spectral analysis method, was established to compute the coupling percentage of multi-channel gastric slow waves recorded using serosal electrodes and electrogastrogram (EGG). Two experiments were performed to validate the method and demonstrate its applications in clinical research. In experiment 1, simultaneous recordings of gastric slow waves were made in five dogs from serosal electrodes and cutaneous electrodes. In experiment 2, four-channel fasting EGGs were made in 10 volunteers for 30 min during waking and 30 min during non-rapid eye movement (REM) sleep. The validation study (experiment 1) showed that the slow wave coupling calculated from the EGGs was correlated with that computed from the serosal recordings. The gastric slow wave coupling percentages detected from both serosal and cutaneous recordings were significantly impaired during vasopressin infusion (6.3 +/- 2.6 vs 62.4 +/- 6.3, P < 0.001 for serosal recordings; 6.7 +/- 3.0 vs 57.2 +/- 2.7, P < 0.001 for cutaneous recordings), and the coupling percentages respectively calculated from serosal and cutaneous recordings were significantly correlated during the baseline recording period (R = 0.922, P < 0.05) and vasopressin infusion period (R = 0.916, P < 0.05). In experiment 2, the gastric slow wave became less coupled when healthy volunteers fell asleep. The percentage of slow wave coupling calculated from the EGGs was 68.2 +/- 17.9% during waking but 41.9 +/- 20.8 during non-REM sleep (P < 0.05). The method developed in this study is reliable for the detection of slow wave uncoupling from multi-channel EGGs. Gastric slow wave coupling is impaired during vasopressin infusion and sleep. These data suggest that this method has potential applications in physiological and clinical studies.
机译:当前的单通道胃电描记法不能检测到胃慢波的耦合或解耦,这对于胃排空至关重要。在这项研究中,建立了一种称为交叉光谱分析方法的新方法,以计算使用浆膜电极和胃电图(EGG)记录的多通道胃慢波的耦合百分比。进行了两个实验以验证该方法并证明其在临床研究中的应用。在实验1中,同时记录了5只来自浆膜电极和皮肤电极的狗的胃慢波。在实验2中,十名志愿者在醒着时30分钟和非快速眼动(REM)睡眠时30分钟制作了四通道的禁食EGG。验证研究(实验1)表明,从EGGs计算出的慢波耦合与从浆膜记录计算出的慢波耦合相关。在输注加压素期间,从浆膜和皮肤记录中检测到的胃慢波耦合百分比均显着受损(浆膜记录的6.3 +/- 2.6与62.4 +/- 6.3,P <0.001; 6.7 +/- 3.0与57.2 +/- 2.7 ,对于皮肤记录,P <0.001),以及分别由浆膜和皮肤记录计算出的耦合百分比在基线记录期间(R = 0.922,P <0.05)和加压素输注期间(R = 0.916,P <0.05)显着相关。 。在实验2中,健康的志愿者入睡后,胃慢波的耦合度降低了。由EGG计算出的慢波耦合百分比在清醒过程中为68.2 +/- 17.9%,但在非REM睡眠过程中为41.9 +/- 20.8(P <0.05)。本研究开发的方法对于检测多通道EGG的慢波解耦是可靠的。加压素输注和睡眠期间胃慢波耦合受损。这些数据表明该方法在生理和临床研究中具有潜在的应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号