首页> 外文期刊>Cancer radiotherapie: journal de la Soci閠?fran鏰ise de radiotherapie oncologique >Development of external quality control protocol for CyberKnife beams dosimetry: Preliminary tests multicentre [Développement d'une procédure de contr?le de qualité externe des faisceaux du CyberKnife: étude préliminaire multicentrique]
【24h】

Development of external quality control protocol for CyberKnife beams dosimetry: Preliminary tests multicentre [Développement d'une procédure de contr?le de qualité externe des faisceaux du CyberKnife: étude préliminaire multicentrique]

机译:射波光束剂量学外部质量控制协议的开发:初步的多中心测试

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: To develop an external quality control procedure for CyberKnife? beams. This work conducted in Nancy, has included a test protocol initially drawn by the medical physicist of Nancy and Lille in collaboration with Equal-Estro Laboratory. Materials and methods: A head and neck anthropomorphic phantom and a water-equivalent homogeneous cubic plastic test-object, so-called "MiniCube", have been used. Powder and solid thermoluminescent dosimeters as well as radiochromic films have been used to perform absolute and relative dose studies, respectively. The comparison between doses calculated by Multiplan treatment planning system and measured doses have been studied in absolute dose. The dose distributions measured with films and treatment planning system calculations have been compared via the gamma function, configured with different tolerance criteria. Results: This work allowed, via solid thermoluminescent dosimeter measurements, verifying the beam reliability with a reproducibility of 1.7%. The absolute dose measured in the phantom irradiated by the seven participating centres has shown an error inferior to the standard tolerance limits (±5%), for most of participating centres. The relative dose measurements performed at Nancy and by the Equal-Estro laboratory allowed defining the most adequate parameters for gamma index (5%/2. mm - with at least 95% of pixels satisfying acceptability criteria: γ. . 1). These parameters should be independent of the film analysis software. Conclusion: This work allowed defining a dosimetric external quality control for CyberKnife? systems, based on a reproducible irradiation plan through measurements performed with thermoluminescent dosimeters and radiochromic films. This protocol should be validated by a new series of measurement and taking into account the lessons of this work.
机译:目的:为Cyber​​Knife开发外部质量控制程序?梁。在南希进行的这项工作包括一个由南希和里尔的医学物理学家与Estro-Estro实验室合作起草的测试协议。材料和方法:已使用头颈部拟人化体模和与水等效的均质立方塑料测试对象,即所谓的“ MiniCube”。粉末和固体热发光剂量计以及放射致变色膜已分别用于进行绝对剂量和相对剂量研究。用绝对剂量研究了通过Multiplan治疗计划系统计算出的剂量与测量剂量之间的比较。通过伽马函数比较了用胶片和治疗计划系统计算得出的剂量分布,并配置了不同的公差标准。结果:这项工作通过固体热发光剂量计测量,以1.7%的可再现性验证了光束的可靠性。对于大多数参与中心,在七个参与中心照射的体模中测量的绝对剂量显示出的误差低于标准公差极限(±5%)。在南希和Equal-Estro实验室进行的相对剂量测量可以定义最合适的伽玛指数参数(5%/ 2。mm-至少95%的像素满足可接受标准:γ。<。1)。这些参数应独立于胶片分析软件。结论:这项工作可以为射波刀定义剂量学外部质量控制?系统是基于可重现的辐射计划,通过使用热发光剂量计和放射致变色膜进行的测量得出。该协议应通过一系列新的测量方法加以验证,并考虑到这项工作的经验教训。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号