In the literature, no consensus exists about optimal irrigation of joints during arthroscopic operations. The goal of this paper is to study the behavior of irrigation systems resulting in the proposal of guidelines for optimal irrigation. To this end, optimal irrigation is defined as the steady state of irrigation of a joint in which a sufficient positive intra-articular pressure and a sufficient flow are maintained. A model of the complete irrigation system was created to schematically elucidate the behavior of pump systems. Additionally, clinical experiments were performed during arthroscopic knee operations in which the pressure at different locations and the irrigation flow were measured. The combination of model prediction and clinical results could well be used to derive guidelines, since the clinical results, which showed considerable variation, were used to verify the model, and the model could be used to explain the typical trends. The main findings are twofold the set pressure is always higher than the intra-articular pressure, and the scope-sheath combination has a significant influence on irrigation control, because of its large restriction. Based on the results, we advice to increase the set pressure during active suction, and to include the sheath-scope combination in the control loop.
机译:在文献中,关于关节镜手术中关节的最佳冲洗尚无共识。本文的目的是研究灌溉系统的行为,从而提出最佳灌溉准则的建议。为此,最佳冲洗被定义为关节的冲洗的稳定状态,在该状态下保持足够的关节内正压和足够的流量。创建了完整灌溉系统的模型,以示意性地阐明泵系统的行为。另外,在关节镜膝关节手术期间进行了临床实验,其中测量了不同位置的压力和冲洗流量。模型预测和临床结果的结合可以很好地用于得出指南,因为临床结果显示出很大的差异,可以用来验证模型,并且该模型可以用来解释典型趋势。主要发现是设定压力始终高于关节内压力的两倍,并且由于其很大的局限性,鞘管组合对灌溉控制有重要影响。根据结果​​,我们建议增加主动吸气期间的设定压力,并将鞘管镜组合纳入控制回路。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号