A new tourniquet system that determines pressures in synchrony with systolic blood pressure in knee surgery.

机译:一种新的止血带系统,可在膝盖手术中与收缩压同步确定压力。

获取原文
获取原文并翻译 | 示例
       

摘要

This study reports the results of the clinical use of a new tourniquet system for surgery related to knee joint that can determine tourniquet pressure in synchrony with systolic blood pressure (SBP). We prospectively applied additional pressure of 100 mmHg based on the SBP recorded just before infiltration of tourniquet in 163 consecutive procedures, which consisted of 73 TKA related surgeries, 42 arthroscopic related surgeries, 20 ACL reconstructed surgeries, 15 ORIF, and 13 other surgeries. After the infiltration, the tourniquet pressure was automatically synchronized with the SBP. The interval of the measurement of blood pressure was every 2.5 min. The average durations of the operation and tourniquet use were 56 (range 6-134) and 56 (range 8-125) min, respectively. The average initial systolic tourniquet pressure was 207 mmHg (range 178-302). The average maximum changes during surgery were 37 mmHg (range 2-102) systolic and 23 mmHg (range 5-67) diastolic. The average maximum systolic pressure changes during 2.5 min interval were 20 mmHg (range 5-97). All cases kept a good operative field without measurable bleeding and there were no postoperative complications.Sixty-eight (42%) cases had a lower intra-operative SBP than the initial value. Since a tourniquet should be applied at the lowest pressure, and for the shortest amount of time possible, the new system appears to be practical and reasonable for maintaining a bloodless surgical field, as compared to conventional tourniquets, which maintain the initial pressure between 300 and 350 mmHg (253 words).
机译:这项研究报告了一种新的止血带系统用于膝关节手术的临床应用结果,该系统可以确定止血带压力与收缩压(SBP)同步。我们基于连续163次手术(包括73例TKA相关手术,42例与关节镜相关的手术,20例ACL重建手术,15例ORIF和13例其他手术)基于止血带浸润前记录的SBP施加了100 mmHg的额外压力。渗透后,止血带压力自动与SBP同步。血压的测量间隔为每2.5分钟一次。手术和止血带的平均使用时间分别为56分钟(6-134)和56分钟(8-125)。平均初始收缩期止血带压力为207 mmHg(范围178-302)。手术期间的平均最大变化为收缩压37 mmHg(范围2-102)和舒张压23 mmHg(范围5-67)。在2.5分钟间隔内平均最大收缩压变化为20 mmHg(范围5-97)。所有病例均保持良好的手术视野,没有可测量的出血,也没有术后并发症.68例(42%)的患者术中SBP低于初始值。由于止血带应在最低压力下并在最短时间内使用,因此与常规止血带相比,新系统对于维持无血外科手术场似乎是实用且合理的,传统止血带将初始压力保持在300至350毫米汞柱(253字)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号