Optimum design of the structure of the electrode for a medical EIT system

机译:医疗EIT系统电极结构的优化设计

获取原文
获取原文并翻译 | 示例
           

摘要

To study the distribution of the sensing field affected by the configuration of the compound electrodes used in the medical EIT system, a finite element model with the condition of the coercive equipotential nodes on the boundary of an object is established. The distribution of the sensing field affected by the width of the compound electrodes is analysed using the simulation method and an optimum design of the structure size of the compound electrode for various excitation strategies is obtained.
机译:为了研究受医疗EIT系统中使用的复合电极的配置影响的感测场的分布,建立了在物体边界上具有矫顽等势节点条件的有限元模型。利用模拟方法分析了受复合电极宽度影响的感应场的分布,并针对各种激励策略获得了复合电极结构尺寸的优化设计。