首页> 外文期刊>Microvascular Research: An International Journal >Variability and repeatability of retinal blood flow measurements using the Canon Laser Blood Flowmeter.
【24h】

Variability and repeatability of retinal blood flow measurements using the Canon Laser Blood Flowmeter.

机译:使用佳能激光血液流量计测量视网膜血流的变异性和可重复性。

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this work was to determine the within-session variability and between-session repeatability of the Canon Laser Blood Flowmeter (CLBF), Model 100, an instrument that permits the noninvasive measurement of retinal blood flow. The CLBF calculates flow in microl/min based on the Poiseuille principle. One eye of 20 normal subjects (mean age 36.5; SD 9.7 years) was randomly chosen. A minimum of five measurements was acquired of a temporal arteriole approximately 1 disc diameter from the optic nerve. Measurements were repeated within a 1-month period. Blood pressure and intraocular pressure were measured. The coefficient of variation (COV) and the coefficient of repeatability (COR) were calculated for each individual. The individual COVs for diameter, velocity, and flow ranged from 0.5 to 6.5% (median 2.0%), 4.8 to 39.7% (median 19.9%), and 4.8 to 37.3% (median 19.3%), respectively. The group mean CORs for diameter, velocity, and flow were 5.2 microm (relative to a mean effect of 104.6 microm),8.8 mm/s (relative to a mean effect of 33.9 mm/s), and 2.6 microl/min (relative to a mean effect of 8.8 microl/min), respectively. The CLBF gave consistent and repeatable measurements of blood flow within retinal arterioles in normal subjects. Given the range of individual variation in the velocity measurement, and thus flow, confidence limits for retinal hemodynamics need to be determined on an individual basis.
机译:这项工作的目的是确定100型佳能激光血液流量计(CLBF)的疗程内变异性和疗程间可重复性,该仪器可无损测量视网膜血流。 CLBF根据Poiseuille原理以微升/分钟计算流量。随机选择20名正常受试者(平均年龄36.5; SD 9.7岁)中的一只眼睛。从视神经获取大约1个椎间盘直径的至少5次测量值。在1个月内重复测量。测量血压和眼内压。计算每个个体的变异系数(COV)和重复性系数(COR)。直径,速度和流量的单个COV分别为0.5%至6.5%(中位数2.0%),4.8%至39.7%(中位数19.9%)和4.8%至37.3%(中位数19.3%)。该组的直径,速度和流量的平均COR为5.2微米(相对于104.6微米的平均效果),8.8毫米/秒(相对于33.9毫米/秒的平均效果)和2.6微升/分钟(相对于32.6毫米/分钟的平均效果)。平均效果分别为8.8微升/分钟)。 CLBF对正常受试者的视网膜小动脉内的血流进行了一致且可重复的测量。给定速度测量中个体变化的范围,从而确定流量,需要根据个体确定视网膜血流动力学的置信度极限。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号