【24h】

High precision comparison between a programmable and a pulse-driven Josephson voltage standard

机译:可编程和脉冲驱动的约瑟夫森电压标准之间的高精度比较

获取原文
获取原文并翻译 | 示例
       

摘要

Over the last 15 years, research in ac Josephson voltage metrology has focused on two fundamentally different systems: the programmable and the pulse-driven Josephson voltage standards (JVSs). This paper reports the first high precision comparison between the two types of JVS. The METAS programmable voltage standard was moved from Switzerland to the Netherlands to be compared with the Dutch pulse-driven system during four days in November 2010. After a careful investigation of the systematic sources of errors, the comparison was made at a frequency of 500 Hz and an rms amplitude of 104 mV. At that level, the voltage difference measured between the fundamental frequency components of the two standards was -0.18 +- 0.26 (mu)V V~(-1) (k velence 2), showing an excellent agreement between the two systems.
机译:在过去的15年中,交流Josephson电压计量学的研究集中在两个根本不同的系统上:可编程和脉冲驱动的Josephson电压标准(JVS)。本文报告了两种JVS之间的第一个高精度比较。 METAS可编程电压标准在2010年11月的四天内从瑞士移至荷兰,与荷兰的脉冲驱动系统进行了比较。在仔细研究了系统性误差源之后,在500 Hz的频率下进行了比较均方根振幅为104 mV。在该水平下,在两个标准的基频分量之间测得的电压差为-0.18 +-0.26μVV〜(-1)(速度2),表明两个系统之间的一致性极好。