摘要

We discuss the treatment of distributed cavity phase, microwave lensing and microwave leakage in the paper by Ovchinnikov and Marra (2011 Metrologia 48 87-100). The paper neglects the potential distributed cavity phase shifts from linear phase gradients and quadrupolar phase variations. Only azimuthally symmetric phase variations were analysed and an incorrect model was used for these. The paper also omits an uncertainty due to microwave lensing, which must be included. Finally, we describe additional measurements that could clarify the model used to analyse the frequency shifts due to microwave leakage.
机译:我们讨论Ovchinnikov和Marra(2011 Metrologia 48 87-100)在论文中对分布式腔相,微波透镜和微波泄漏的处理。本文忽略了线性相位梯度和四极相位变化引起的潜在的腔体相移。仅分析了方位角对称的相位变化,并为此使用了错误的模型。本文还忽略了由于微波透镜引起的不确定性,必须将其包括在内。最后,我们描述了其他度量,这些度量可以阐明用于分析由于微波泄漏引起的频率偏移的模型。