Influence of the orientation of the Osstell transducer during measurement of dental implant stability using resonance frequency analysis: a numerical approach.

机译:Osstell换能器定向在使用共振频率分析的牙科植入物稳定性测量过程中的影响:一种数值方法。

获取原文
获取原文并翻译 | 示例
           

摘要

Stability of dental implants is measured by means of the Osstell device using a method of resonance frequency analysis. The aim of the study was to evaluate the role of direction-dependence of the Osstell transducer. For this purpose, a set of parametrical finite element analyses has been used. When rotating the transducer around the vertical axis the crucial phenomenon is the change in behavior of the system, which occurs in positions between 30 degrees and 80 degrees to the long axis of the alveolar crest. It seems from the presented results that, when measuring approximately perpendicularly to this long axis, the deviation from the ideal perpendicular position must not exceed 30 degrees. In this case, the first resonance frequency is recorded. When measuring in the approximately parallel position to the long axis of the alveolar crest, the second resonance frequency is recorded. The deviation from the ideal parallel position must not exceed 10 degrees. These values are not comparable with the previous ones.
机译:牙植入物的稳定性通过Osstell设备使用共振频率分析方法进行测量。该研究的目的是评估Osstell换能器的方向依赖性作用。为此,已使用了一组参数化有限元分析。当围绕垂直轴旋转换能器时,关键现象是系统行为的变化,这种变化发生在与牙槽c长轴成30度到80度之间的位置。从给出的结果看来,当近似垂直于该长轴进行测量时,与理想垂直位置的偏差不得超过30度。在这种情况下,记录第一共振频率。当在与牙槽the的长轴大致平行的位置进行测量时,将记录第二共振频率。与理想平行位置的偏差不得超过10度。这些值与以前的值不具有可比性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号