Purpose: A unique capability of the CyberKnife system is dynamic target tracking. However, not all patients are eligible for this approach. Rather, their tumors are tracked statically using the vertebral column for alignment. When using static tracking, the internal target volume (ITV) is delineated on the four-dimensional (4D) CT scan and an additional margin is added to account for setup uncertainty [planning target volume (PTV)]. Treatment margins are difficult to estimate due to unpredictable variations in tumor motion and respiratory pattern during the course of treatment. The inability to track the target and detect changes in respiratory characteristics might result in geographic misses and local tumor recurrences. The purpose of this study is to develop a method to evaluate the adequacy of ITV-to-PTV margins for patients treated in this manner.
机译:目的:数码刀系统的独特功能是动态目标跟踪。但是,并非所有患者都有资格使用此方法。而是使用椎骨柱对它们的肿瘤进行静态跟踪以进行对准。当使用静态跟踪时,内部目标体积(ITV)在四维(4D)CT扫描上划定,并添加了额外的余量以解决设置不确定性[计划目标体积(PTV)]。由于在治疗过程中肿瘤运动和呼吸模式的不可预测的变化,因此难以估计治疗余量。无法跟踪目标并检测呼吸特征的变化可能导致地理遗漏和局部肿瘤复发。这项研究的目的是开发一种方法,以评估以这种方式治疗的患者的ITV到PTV边缘的充分性。