A cascaded model of spectral distortions due to spectral response effects and pulse piSeup effects in a photon-counting x-ray detector for CT

机译:用于CT的光子计数X射线探测器中由光谱响应效应和脉冲piSeup效应引起的光谱畸变的级联模型

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: Energy discriminating, photon-counting detectors (PCDs) are an emerging technology for computed tomography (CT) with various potential benefits for clinical CT. The photon energies measured by PCDs can be distorted due to the interactions of a photon with the detector and the interaction of multiple coincident photons. These effects result in distorted recorded x-ray spectra which may lead to artifacts in reconstructed CT images and inaccuracies in tissue identification. Model-based compensation techniques have the potential to account for the distortion effects. This approach requires only a small number of parameters and is applicable to a wide range of spectra and count rates, but it needs an accurate model of the spectral distortions occurring in PCDs. The purpose of this study was to develop a model of those spectral distortions and to evaluate the model using a PCD (model DXMCT-1; DxRay, Inc., Northridge, CA) and various x-ray spectra in a wide range of count rates. Methods: The authors hypothesize that the complex phenomena of spectral distortions can be modeled by: (1) separating them into count-rate independent factors that we call the spectral response effects (SRE), and count-rate dependent factors that we call the pulse pileup effects (PPE), (2) developing separate models for SRE and PPE, and (3) cascading the SRE and PPE models into a combined SRE+PPE model that describes PCD distortions at both low and high count rates. The SRE model describes the probability distribution of the recorded spectrum, with a photo peak and a continuum tail, given the incident photon energy. Model parameters were obtained from calibration measurements with three radioisotopes and then interpolated linearly for other energies. The PPE model used was developed in the authors' previous work [K. Taguchi et al, "Modeling the performance of a photon counting x-ray detector for CT: Energy response and pulse pileup effects," Med. Phys. 38(2), 1089-1102 (2011)]. The agreement between the x-ray spectra calculated by the cascaded SRE+PPE model and the measured spectra was evaluated for various levels of deadtime loss ratios (DLR) and incident spectral shapes, realized using different attenuators, in terms of the weighted coefficient of variation (COVw), i.e., the root mean square difference weighted by the statistical errors of the data and divided by the mean. Results: At low count rates, when DLR < 10%, the distorted spectra measured by the DXMCT-1 were in agreement with those calculated by SRE only, with COVw's less than 4%. At higher count rates, the measured spectra were also in agreement with the ones calculated by the cascaded SRE+PPE model; with PMMA as attenuator, COV_W was 5.6% at a DLR of 22% and as small as 6.7% for a DLR as high as 55%. Conclusions: The x-ray spectra calculated by the proposed model agreed with the measured spectra over a wide range of count rates and spectral shapes. The SRE model predicted the distorted, recorded spectra with low count rates over various types and thicknesses of attenuators. The study also validated the hypothesis that the complex spectral distortions in a PCD can be adequately modeled by cascading the count-rate independent SRE and the count-rate dependent PPE.
机译:目的:能量辨别,光子计数检测器(PCD)是计算机断层摄影(CT)的新兴技术,具有临床CT的各种潜在好处。由于光子与检测器的相互作用以及多个重合光子的相互作用,PCD测得的光子能量可能会失真。这些影响导致记录的X射线光谱失真,这可能导致重建的CT图像中出现伪影,并且组织识别不准确。基于模型的补偿技术具有解决失真影响的潜力。这种方法仅需要少量参数,并且适用于广泛的光谱和计数率,但它需要PCD中发生的光谱失真的精确模型。这项研究的目的是建立一个光谱畸变的模型,并使用PCD(DXMCT-1型; DxRay,Inc.,Northridge,CA)和各种计数率下的各种X射线光谱对模型进行评估。 。方法:作者假设可以通过以下方式对复杂的频谱失真现象进行建模:(1)将其分为计数率无关的因素(我们称为频谱响应效应,SRE)和计数率相关的因素(称为脉冲)。堆积效应(PPE),(2)为SRE和PPE开发单独的模型,以及(3)将SRE和PPE模型层叠为组合的SRE + PPE模型,该模型描述低计数率和高计数率的PCD失真。在给定入射光子能量的情况下,SRE模型描述了记录光谱的概率分布,具有光峰和连续峰。模型参数是通过使用三个放射性同位素的校准测量获得的,然后线性内插其他能量。使用的PPE模型是作者先前的工作开发的[K. Taguchi等人,“为CT模拟光子计数X射线检测器的性能:能量响应和脉冲堆积效应”,医学杂志。物理38(2),1089-1102(2011)]。通过级联SRE + PPE模型计算的X射线光谱与测得的光谱之间的一致性,评估了不同水平的空载时间损耗比(DLR)和入射光谱形状,并使用了不同的衰减器根据加权变异系数(COVw),即均方根差乘以数据的统计误差然后除以均值。结果:在低计数率下,当DLR <10%时,由DXMCT-1测量的失真光谱与仅由SRE计算的光谱一致,COVw小于4%。在更高的计数率下,测得的光谱也与级联的SRE + PPE模型所计算的光谱一致。如果使用PMMA作为衰减器,则DLR为22%时COV_W为5.6%,而DLR高达55%时COV_W仅为6.7%。结论:所提出的模型计算的X射线光谱与在宽范围的计数率和光谱形状上测得的光谱一致。 SRE模型以各种计数和厚度的衰减器的低计数率预测失真的记录光谱。该研究还证实了这样的假设:通过将独立于计数率的SRE和依赖于计数率的PPE级联,可以适当地建模PCD中的复杂频谱失真。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号