Modeling photon-counting detectors for x-ray CT: spectral response and pulse-pileup effects and evaluation using real data

机译:对用于X射线CT的光子计数检测器建模:光谱响应和脉冲堆积效应,并使用实际数据进行评估

获取原文

摘要

Spectral computed tomography (CT) with photon-counting detectors (PCDs) has the potential to substantially advance diagnostic CT imaging by reducing image noise and dose to the patient, by improving contrast and tissue specificity, and by enabling molecular and functional imaging. The current PCD technology, however, is limited by two main factors: imperfect energy measurement (spectral response effects, SRE) and limited count rate (pulse-pileup effects, PPE, due to detector dead-times). The overall goal of our research is to develop compensation algorithms for these sources of data distortions, to demonstrate that PCDs are suitable for clinical CT, and to identify key clinical applications for spectral CT with PCDs. We have already developed an iterative compensation scheme that includes a forward projection model of the imaging chain and that can compensate for either SRE or PPE distortions separately by maximizing a penalized log-likelihood function. In this paper we describe the evaluation of a combined, cascaded SRE and PPE model for PCDs and compare the models to data acquired with an experimental table-top PCD-CT system. The separation into count-rate independent effects (SRE) and count-rate dependent effects (PPE) allows cascading the forward model. First, the SRE model is evaluated using low count rates. Then the PPE model is cascaded and the combined SRE+PPE model is evaluated. Several different attenuators were used, including K-edge materials and the models and data were compared for various count-rate conditions.
机译:带有光子计数检测器(PCD)的光谱计算机断层扫描(CT)通过降低图像噪声和对患者的剂量,改善对比度和组织特异性以及实现分子和功能成像,具有显着提高诊断CT成像的潜力。但是,当前的PCD技术受到两个主要因素的限制:不完善的能量测量(光谱响应效应,SRE)和有限的计数率(脉冲堆积效应,PPE,由于检测器死区时间)。我们研究的总体目标是针对这些数据失真源开发补偿算法,以证明PCD适用于临床CT,并确定具有PCD的光谱CT的关键临床应用。我们已经开发出一种迭代补偿方案,该方案包括成像链的前向投影模型,并且可以通过最大化惩罚对数似然函数来分别补偿SRE或PPE失真。在本文中,我们描述了PCD的组合,级联SRE和PPE模型的评估,并将该模型与通过实验台式PCD-CT系统获取的数据进行比较。将计数速率无关效应(SRE)和计数速率依赖性效应(PPE)分开可以级联正向模型。首先,使用低计数率评估SRE模型。然后将PPE模型级联,并评估组合的SRE + PPE模型。使用了几种不同的衰减器,包括K边缘材料,并对各种计数率条件下的模型和数据进行了比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号