Evaluation of delivered monitor unit accuracy of gated step-and-shoot IMRT using a two-dimensional detector array.

机译:使用二维检测器阵列评估门控步进式IMRT交付的监控器精度。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To overcome the problem of organ motion in intensity-modulated radiation therapy (IMRT), gated IMRT is often used for the treatment of lung cancer. In this study, the authors investigated the accuracy of the delivered monitor units (MUs) from each segment during gated IMRT using a two-dimensional detector array for user-specific verification purpose. METHODS: The authors planned a 6 MV photon, seven-port step-and-shoot lung IMRT delivery. The respiration signals for gated IMRT delivery were obtained from the one-dimensional moving phantom using the real-time position management (RPM) system (Varian Medical Systems, Palo Alto, CA). The beams were delivered using a Clinac iX (Varian Medical Systems, Palo Alto, CA) with the Millennium 120 MLC. The MatriXX (IBA Dosimetry GmbH, Germany) was validated through consistency and reproducibility tests as well as comparison with measurements from a Farmer-type ion chamber. The authors delivered beams with varying dose rates and duty cycles and analyzed the MatriXX data to evaluate MU delivery accuracy. RESULTS: There was quite good agreement between the planned segment MUs and the MUs computed from the MatriXX within +/- 2% error. The beam-on times computed from the MatriXX data were almost identical for all cases, and they matched well with the RPM beam-on and beam-off signals. A slight difference was observed between them, but it was less than 40 ms. The gated IMRT delivery demonstrated an MU delivery accuracy that was equivalent to ungated IMRT, and the delivered MUs with a gating signal agreed with the planned MUs within +/- 0.5 MU regardless of dose rate and duty cycle. CONCLUSIONS: The authors can conclude that gated IMRT is able to deliver an accurate dose to a patient during a procedure. The authors believe that the methodology and results can be transferred to other vendors' devices, particularly those that do not provide MLC log data for a verification purpose.
机译:目的:为克服强度调制放射治疗(IMRT)中器官运动的问题,门控IMRT通常用于治疗肺癌。在这项研究中,作者使用二维检测器阵列针对用户特定的验证目的,研究了门控IMRT期间每个段的已交付监控单元(MU)的准确性。方法:作者计划了6 MV光子,七端口分步式肺部IMRT递送。使用实时位置管理(RPM)系统(Varian Medical Systems,Palo Alto,CA)从一维运动体模中获得用于门控IMRT递送的呼吸信号。使用Clinac iX(加利福尼亚州帕洛阿尔托市的瓦里安医疗系统公司)和Millennium 120 MLC传输光束。 MatriXX(IBA Dosimetry GmbH,德国)已通过一致性和可重复性测试以及与Farmer型离子室测量结果的比较进行了验证。作者提供了具有不同剂量率和占空比的光束,并分析了MatriXX数据以评估MU的传送精度。结果:计划的段MU和从MatriXX计算的MU之间有很好的一致性,误差在+/- 2%之内。从MatriXX数据计算出的束通时间在所有情况下几乎相同,并且它们与RPM束通和束离信号非常匹配。他们之间观察到细微的差别,但不到40毫秒。门控IMRT交付证明了MU交付的准确性与非门控IMRT相当,无论门诊剂量率和占空比如何,门控信号与计划MU一致的交付MU均在+/- 0.5 MU之内。结论:作者可以得出结论,门控IMRT能够在手术过程中向患者提供准确的剂量。作者认为,该方法和结果可以转移到其他供应商的设备上,尤其是那些不提供MLC日志数据进行验证的设备。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号