Image artifacts, caused by a temporally delayed response of a computed tomography (CT) detector, were investigated. To study its consequences, a computer model for a standard third generation CT scanner was devised and simulations were carried out. Resulting image artifacts were studied for various voxel-based and mathematical phantoms using three time constants, which characterize the delayed signal responses of different detector configurations. Furthermore, a theory was developed to understand temporal artifacts in reconstructed images. A filter function was derived, which compensates for temporal artifacts. For a given phantom, simulations and theory demonstrate that artifacts scale with the time constant, characteristic for the detector, and the angular speed of the scanner.
机译:研究了由计算机断层扫描(CT)检测器的时间延迟响应引起的图像伪影。为了研究其后果,设计了用于第三代标准CT扫描仪的计算机模型,并进行了仿真。使用三个时间常数对各种基于体素和数学模型的图像伪影进行了研究,这些常数表征了不同检测器配置的延迟信号响应。此外,发展了一种理论来理解重建图像中的时间伪像。导出了一个滤波函数,该函数补偿了时间伪像。对于给定的幻像,仿真和理论证明了伪像随时间常数,检测器的特性以及扫描仪的角速度而缩放。