首页> 外文会议>IEEE International Symposium on Biomedical Imaging >Motion artifacts in kidney stone imaging using dual-energy CT. A phantom study comparing single-source and dual-source scanners
【24h】

Motion artifacts in kidney stone imaging using dual-energy CT. A phantom study comparing single-source and dual-source scanners

机译:使用双能CT在肾结石成像中的运动伪影。对单源和双源扫描仪进行比较的幻像研究

获取原文

摘要

Computed tomography (CT) has been established as the method of choice for imaging kidney stones with high sensitivity and accuracy. Further, dual-energy CT (DECT) showed the capability of differentiating uric-acid (UA) from non-UA stones with 90-100% accuracy. DECT depends on acquiring two images at different energy levels, from which a material-specific color-coded image is generated. With the invention of dual-source (DS) scanners, both low- and high-energy images are acquired simultaneously. However, DECT can also be performed by sequential acquisition of both images on single-source (SS) scanners. Although motion artifacts generally result in distorted anatomy and blurring, in stone imaging they may lead to stone misclassification. In this study, we investigate the effects of motion artifacts on stone classification using both SS and DS DECT. The results show that stone misclassification may be significantly higher in SS DECT, especially when multiple stones with different types lie in close proximity.
机译:已经建立了计算机断层扫描(CT)作为对肾结石进行高灵敏度和准确度成像的首选方法。此外,双能CT(DECT)显示出以90-100%的准确度区分尿酸(UA)与非UA结石的能力。 DECT取决于获取处于不同能级的两个图像,从中生成特定于材料的颜色编码图像。利用双源(DS)扫描仪的发明,可以同时获取低能和高能图像。但是,也可以通过在单源(SS)扫描仪上顺序采集两个图像来执行DECT。尽管运动伪影通常会导致解剖结构变形和模糊,但在结石成像中,它们可能会导致结石分类错误。在这项研究中,我们使用SS和DS DECT研究运动伪影对石材分类的影响。结果表明,SS DECT中的宝石分类错误可能会更高,尤其是当多个不同类型的宝石紧邻时。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号