Implementation and validation of three automated methods for measuring ultrasound maximum depth of penetration: application to ultrasound quality control.

机译:三种自动测量超声最大穿透深度的方法的实施和验证:在超声质量控制中的应用。

获取原文
获取原文并翻译 | 示例

摘要

Systematic measurements of maximum depth of penetration (DOP) of ultrasound (US) scanners are essential for quality control (QC). Conventionally, DOP measurements are performed visually and as such they could be affected by various external factors, scanner control settings, and operator related errors. Automated methods should be free of the issues associated with interoperator dependence and are an attractive alternative to the visual DOP measurements. We implement and test three automated methods for measuring DOP. The methods base their measurements on signal to noise (SNR) analysis of uniform US phantom images. Two of the methods use pairs of phantom images. The third one uses a single phantom image and an "in-air" image. The validation tests included precision, sensitivity, repeatability, and usability in routine QC application. Methods based on pairs of phantom images measure the DOP with precision +/-0.2 cm or better. Precision of the single phantom image method is +/-0.05 cm, and that method is also the most sensitive of the three. All three methods are demonstrated to be repeatable among different users. Since the images for the DOP computation are collected free-hand the sensitivity to hand-transducer motion during image acquisition was also tested. Unlike the single-phantom-image based method, the methods using image pairs were found to be very sensitive to transducer motion and therefore less convenient for clinical QC applications. In conclusion, the single-phantom-image method is best suited for routine QC in a real-life clinical practice.
机译:超声(US)扫描仪的最大穿透深度(DOP)的系统测量对于质量控制(QC)至关重要。通常,DOP测量是在视觉上执行的,因此,它们可能会受到各种外部因素,扫描仪控制设置以及操作员相关错误的影响。自动化方法应避免与操作员依赖性相关的问题,并且是可视DOP测量的有吸引力的替代方法。我们实施并测试了三种测量DOP的自动化方法。这些方法的测量基于统一的美国幻像图像的信噪比(SNR)分析。其中两种方法使用成对的幻像图像。第三个使用单个幻像图像和“空中”图像。验证测试包括常规QC应用程序中的精度,灵敏度,可重复性和可用性。基于幻像对的方法可以以+/- 0.2 cm或更高的精度测量DOP。单个幻影图像方法的精度为+/- 0.05 cm,该方法也是这三种方法中最敏感的。这三种方法被证明在不同用户之间是可重复的。由于用于DOP计算的图像是徒手收集的,因此还测试了在图像采集过程中对换能器运动的敏感性。与基于单幻像图像的方法不同,发现使用图像对的方法对换能器运动非常敏感,因此对于临床质量控制应用不太方便。总之,单幻影方法最适合现实生活中的常规QC。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号