...
首页> 外文期刊>Medical Physics >Evaluation of clip localization for different kilovoltage imaging modalities as applied to partial breast irradiation setup.
【24h】

Evaluation of clip localization for different kilovoltage imaging modalities as applied to partial breast irradiation setup.

机译:应用于局部乳房照射设置的不同千伏成像模态的夹具定位评估。

获取原文
获取原文并翻译 | 示例

摘要

Surgical clip localization and image quality were evaluated for different types of kilovoltage cone beam imaging modalities as applied to partial breast irradiation (PBI) setup. These modalities included (i) clinically available radiographs and cone beam CT (CB-CT) and (ii) various alternative modalities based on partial/sparse/truncated CB-CT. An anthropomorphic torso-breast phantom with surgical clips was used for the imaging studies. The torso phantom had artificial lungs, and the attached breast phantom was a mammographic phantom with realistic shape and tissue inhomogeneities. Three types of clips of variable size were used in two orthogonal orientations to assess their in-/cross-plane characteristics for image-guided setup of the torso-breast phantom in supine position. All studies were performed with the Varian on-board imaging (OBI, Varian) system. CT reconstructions were calculated with the standard Feldkamp-Davis-Kress algorithm. First, the radiographs were studied for a wide range of viewingangles to characterize image quality for various types of body anatomy in the foreground/background of the clips. Next, image reconstruction quality was evaluated for partial/sparse/truncated CB-CT. Since these modalities led to reconstructions with strong artifacts due to insufficient input data, a knowledge-based CT reconstruction method was also tested. In this method, the input data to the reconstruction algorithm were modified by combining complementary data sets selected from the treatment and reference projections. Different partial/sparse/truncated CB-CT scan types were studied depending on the total are angle, angular increment between the consequent views (CT projections), orientation of the arc center with respect to the imaged breast and chest wall, and imaging field size. The central angles of the viewing arcs were either tangential or orthogonal to the chest wall. Several offset positions of the phantom with respect to the reference position were studied. The acquired and reconstructed image data sets were analyzed using home-built software focusing on the ability to localize clips in 3D. Streaking and leakage reconstruction artifacts and spatial distortions of breast surface were analyzed as well. Advantages and disadvantages of each kilovoltage CB imaging modality as applied to partial breast setup evaluation based on clips are presented. Because clips were found to be difficult to recognize in radiographs, 3D reconstructions were preferred. Even though it was possible to localize clips with about +/-1 mm accuracy based on reconstructions for short arcs of 40 degrees and incremental angle up to about 5 degrees, identification of clips in such reconstructions is difficult. Reconstructions obtained for arcs of as low as 80 degrees and incremental angle of as high as 3 degrees were suggested for easier clip identification. For more severely undersampled data, iterative CB-CT reconstruction is recommended to decrease the artifacts.
机译:对于应用于部分乳房照射(PBI)设置的不同类型的千伏锥束成像模式,评估了手术夹的定位和图像质量。这些模式包括(i)临床上可获得的X射线照片和锥束CT(CB-CT),以及(ii)基于部分/稀疏/截断CB-CT的各种替代模式。带有手术夹的拟人躯干-乳房幻影用于成像研究。躯干体模具有人工肺,附着的乳房体模是具有逼真的形状和组织不均匀性的乳腺摄影体模。在两个正交的方向上使用了三种类型的可变大小的夹子,以评估它们在仰卧位置上的躯干-乳房幻影的图像引导设置中的/跨平面特性。所有研究均使用Varian车载成像(OBI,Varian)系统进行。使用标准的Feldkamp-Davis-Kress算法计算CT重建。首先,研究了X射线照片的各种视角,以表征剪辑前景/背景中各种类型的人体解剖结构的图像质量。接下来,针对部分/稀疏/截断的CB-CT评估图像重建质量。由于这些模态由于输入数据不足而导致使用强伪影进行重建,因此还测试了基于知识的CT重建方法。在这种方法中,通过组合从治疗和参考投影中选择的补充数据集来修改重建算法的输入数据。研究了不同的部分/稀疏/截断的CB-CT扫描类型,具体取决于角度,结果视图之间的角度增量(CT投影),相对于成像的胸部和胸部壁的弧中心方向以及成像场大小。观察弧的中心角与胸壁成切线或正交。研究了幻影相对于参考位置的几个偏移位置。使用专注于在3D中定位剪辑的功能的自制软件分析获取和重建的图像数据集。还分析了条纹和泄漏重建伪影以及乳房表面的空间变形。提出了每种千伏CB成像方式的优缺点,这些方式适用于基于片段的部分乳房设置评估。由于发现在射线照片中难以识别剪辑,因此首选3D重建。即使有可能基于40度短弧和最多约5度的增量角度的重建,以大约+/- 1 mm的精度定位剪辑,但在这种重建中很难识别剪辑。建议对低至80度的圆弧和递增至3度的增量角进行重建,以便更容易地识别剪辑。对于更严重的欠采样数据,建议使用迭代CB-CT重建以减少伪像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号