首页> 外文期刊>Medical Physics >Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device.
【24h】

Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device.

机译:非晶硅电子门禁成像设备的离轴剂量响应特性。

获取原文
获取原文并翻译 | 示例
       

摘要

Amorphous silicon (a-Si) electronic portal imaging devices (EPIDs) have typically been calibrated to dose at central axis (CAX). Division of acquired images by the flood-field (FF) image that corrects for pixel sensitivity variation as well as open field energy-dependent off-axis response variation should result in a flat EPID response over the entire matrix for the same field size. While the beam profile can be reintroduced to the image by an additional correction matrix, the CAX EPID response to dose calibration factor is assumed to apply to all pixels in the detector. The aim of this work was to investigate the dose response of the Varian aS500 amorphous silicon detector across the entire detector area. First it was established that the EPID response across the panel became stable (within approximately 0.2%) for MU settings greater than approximately 200 MU. The EPID was then FF calibrated with a high MU setting of approximately 400 for all subsequent experiments. Whole detector images with varying MU settings from 2-500 were then acquired for two dose rates (300 and 600 MU/min) for 6 MV photons for two EPIDs. The FF corrected EPID response was approximately flat or uniform across the detector for greater than 100 MU delivered (within 0.5%). However, the off-axis EPID response was greater than the CAX response for small MU irradiations, giving a raised EPID profile. Up to 5% increase in response at 20 cm off-axis compared to CAX was found for very small MU settings for one EPID, while it was within 2% for the second (newer) EPID. Off-axis response nonuniformities attributed to detector damage were also found for the older EPID. Similar results were obtained with the EPID at 18 MV energy and operating in asynchronous mode (acquisition not synchronized with beam pulses), however the profiles were flatter and more irregular for the small MU irradiations. By moving the detector laterally and repeating the experiments, the increase in response off-axis was found to depend on the pixel position relative to the beam CAX. When the beam was heavily filtered by a phantom the off-axis response variation was reduced markedly to within 0.5% for all MU settings. Independent measurements of off-axis point doses with ion chamber did not show any change in off-axis factor with MUs. Measurements of beam quality (TMR20-10) for MU settings of 2, 5, and 100 at central axis and at 15 cm off-axis could not explain the effect. The response change is unlikely to be significant for clinical IMRT verification with this imaging/acclerator system where MUs are of the order of 100-300, provided the detector does not exhibit radiation damage artifacts.
机译:非晶硅(a-Si)电子门禁成像设备(EPID)通常已校准为在中心轴(CAX)处具有剂量。泛光场(FF)图像对获取的图像进行的划分可校正像素灵敏度变化以及与开放场能量有关的离轴响应变化,对于相同的场大小,应在整个矩阵上产生平坦的EPID响应。虽然可以通过附加的校正矩阵将光束轮廓重新引入图像,但假定对剂量校准因子的CAX EPID响应适用于检测器中的所有像素。这项工作的目的是研究Varian aS500非晶硅探测器在整个探测器区域内的剂量响应。首先,已经确定,对于大于大约200 MU的MU设置,整个面板上的EPID响应变得稳定(在大约0.2%之内)。然后将EPID用大约400的高MU设置进行FF校准,以用于所有后续实验。然后,针对两个EPID的6 MV光子,以两种剂量率(300和600 MU / min)获得具有从2-500变化的MU设置的整个探测器图像。 FF校正的EPID响应在整个检测器上大致平坦或均匀,超过100 MU的输出(在0.5%以内)。但是,对于小MU辐照,离轴EPID响应大于CAX响应,从而提高了EPID曲线。对于一个EPID的非常小的MU设置,与CAX相比,在离轴20 cm处的响应最多增加5%,而对于第二个(较新的)EPID,响应距离在2%之内。对于较旧的EPID,还发现了归因于检测器损坏的离轴响应不均匀性。在18 MV能量下以EPID并以异步模式运行(采集与光束脉冲不同步)时,获得了相似的结果,但是对于小MU辐照而言,轮廓更平坦且更不规则。通过横向移动检测器并重复实验,发现离轴响应的增加取决于像素相对于光束CAX的位置。当光束被幻影严重过滤时,对于所有MU设置,离轴响应变化都显着降低至0.5%以内。用离子室对轴外点剂量的独立测量未显示MU对轴外因子的任何变化。在中心轴和离轴15 cm处的MU设置为2、5和100时,光束质量(TMR20-10)的测量无法解释这种影响。对于这种成像/加速器系统,如果MU的数量在100-300之间,则响应变化对于临床IMRT验证不太可能是重要的,前提是检测器没有辐射损伤伪影。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号