In this study we investigate the effect of detector size in the dosimetry of small fields and steep dose gradients with a particular emphasis on IMRT measurements. Comparisons of calculated and measured cross-profiles and absolute dose values of IMRT treatment plans are presented. As a consequence of the finite size of the detector that was used for the commissioning of the IMRT tool, local discrepancies of more than 10% are found between calculated cross-profiles of intensity modulated beams and intensity modulated profiles measured with film. Absolute dose measurements of intensity modulated fields with a 0.6 cm3 Farmer chamber show significant differences of more than 6% between calculated and measured dose values at the isocenter of an IMRT treatment plan. Differences of not more than 2% are found in the same experiment for dose values measured with a 0.015 cm3 pinpoint ion chamber. A method to correct for the spatial response of finite-sized detectors and to obtain the real Output factor measurements are performed with different detectors and are presented as a function of detector size for a 1 x 1 cm2 field. Because of its high spatial resolution and water equivalence, a diamond detector is found to be suitable as an alternative to other detectors used for small field dosimetry as there are photographic and photochromic film, TLDs, or water-equivalent scintillation detectors.
机译:在这项研究中,我们研究了检测器尺寸对小场和陡峭剂量梯度的剂量测定的影响,尤其着重于IMRT测量。给出了IMRT治疗计划的计算和测量的交叉曲线和绝对剂量值的比较。由于用于IMRT工具调试的检测器尺寸有限,因此在计算出的强度调制光束的交叉轮廓与用胶片测量的强度调制轮廓之间发现了超过10%的局部差异。用0.6 cm3法默室进行强度调制的场的绝对剂量测量结果显示,在IMRT治疗计划的等中心点,计算出的剂量值和测量到的剂量值之间的显着差异超过6%。在同一实验中,使用0.015 cm3精确离子室测量的剂量值差异不超过2%。一种校正有限尺寸探测器的空间响应并获得实际输出因子测量值的方法是使用不同的探测器执行的,并且针对1 x 1 cm2的场,它是探测器尺寸的函数。由于它具有较高的空间分辨率和水等效性,因此发现它具有照相和光致变色胶片,TLD或与水等效的闪烁检测器,因此,它适合作为小剂量法其他检测器的替代产品。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号