首页> 外文期刊>Medical Physics >Online 3D EPID-based dose verification: Proof of concept
【24h】

Online 3D EPID-based dose verification: Proof of concept

机译:基于在线3D EPID的剂量验证:概念验证

获取原文
获取原文并翻译 | 示例

摘要

Purpose: Delivery errors during radiotherapy may lead to medical harm and reduced life expectancy for patients. Such serious incidents can be avoided by performing dose verification online, i.e., while the patient is being irradiated, creating the possibility of halting the linac in case of a large overdosage or underdosage. The offline EPID-based 3D in vivo dosimetry system clinically employed at our institute is in principle suited for online treatment verification, provided the system is able to complete 3D dose reconstruction and verification within 420 ms, the present acquisition time of a single EPID frame. It is the aim of this study to show that our EPID-based dosimetry system can be made fast enough to achieve online 3D in vivo dose verification.
机译:目的:放疗过程中的传送错误可能会导致医疗伤害并缩短患者的预期寿命。可以通过在线进行剂量验证(即在对患者进行辐照时)来避免此类严重事件,从而在大剂量或低剂量情况下停止直线加速器。我们研究所临床上使用的基于离线EPID的3D体内剂量测定系统原则上适用于在线治疗验证,只要该系统能够在420毫秒(即单个EPID帧的当前采集时间)内完成3D剂量重建和验证。这项研究的目的是表明我们的基于EPID的剂量测定系统可以足够快地实现在线3D体内剂量验证。