Transmission of broad W/Rh and W/Al (target/filter) x-ray beams operated at 25-49 kVp through common shielding materials

机译:在25-49 kVp下工作的宽W / Rh和W / Al(目标/滤光片)X射线束通过普通屏蔽材料的透射

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To provide transmission data for broad 25-39 kVp (kilovolt peak) W/Rh and 25-49 kVp W/Al (target/filter, W-tungsten, Rh-rhodium, and Al-aluminum) x-ray beams through common shielding materials, such as lead, concrete, gypsum wallboard, wood, steel, and plate glass. Methods: The unfiltered W-target x-ray spectra measured on a Selenia Dimensions system (Hologic Inc., Bedford, MA) set at 20-49 kVp were, respectively, filtered using 50-μm Rh and 700-μm Al, and were subsequently used for Monte Carlo calculations. The transmission of broad x-ray beams through shielding materials was simulated using Geant4 low energy electromagnetic physics package with photon- and electron-processes above 250 eV, including photoelectric effect, Compton scattering, and Rayleigh scattering. The calculated transmission data were fitted using Archer equation with a robust fitting algorithm. Results: The transmission of broad x-ray beams through the above-mentioned shielding materials was calculated down to about 10 -5 for 25-39 kVp W/Rh and 25-49 kVp W/Al. The fitted results of α, β, and γ in Archer equation were provided. The α values of kVp 40 were approximately consistent with those of NCRP Report No. 147. Conclusions: These data provide inputs for the shielding designs of x-ray imaging facilities with W-anode x-ray beams, such as from Selenia Dimensions.
机译:目的:提供宽的25-39 kVp(千伏峰值)W / Rh和25-49 kVp W / Al(目标/滤光片,钨-钨,铑-铑和铝-铝)x射线束的透射数据常见的屏蔽材料,例如铅,混凝土,石膏墙板,木材,钢和平板玻璃。方法:分别在设置为20-49 kVp的Selenia Dimensions系统(Hologic Inc.,贝德福德,马萨诸塞州)上测量未过滤的W靶x射线光谱,分别使用50μmRh和700μmAl过滤,并随后用于蒙特卡洛计算。使用Geant4低能电磁物理封装对250xV以上的光子和电子过程进行了模拟,分析了宽X射线束通过屏蔽材料的透射情况,包括光电效应,康普顿散射和瑞利散射。使用Archer方程和鲁棒拟合算法对计算出的传输数据进行拟合。结果:对于25-39 kVp W / Rh和25-49 kVp W / Al,通过上述屏蔽材料的宽X射线束的透射率可低至约10 -5。提供了Archer方程中α,β和γ的拟合结果。 kVp 40的α值与NCRP报告147的α值大致一致。结论:这些数据为带有W阳极X射线束的X射线成像设备的屏蔽设计提供了输入,例如Selenia Dimensions。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号