A dosimetric study of a new 125I seed for permanent prostate implant, the Symmetra 1251 Seed model I25.S06, has been undertaken utilizing Monte Carlo photon transport calculations. All dosimetric quantities recommended by the AAPM Task Group 43 (TG-43) report have been calculated. Quantities determined are dose rate constant, radial dose function, anisotropy function, anisotropy factor, and anisotropy constant. The recently (January 1999) revised NIST (National Institute of Standards and Technology) 1251 standard for air kerma strength calibration was taken into account as well as updated interaction cross-section data. Calculations were done for the competing model 6702 source for the purpose of comparison. The calculated dose-rate constants for the two seeds are 1.010 and 1.016 cGyh(-1)U(-1) for the Symmetra and model 6702 seeds, respectively. The latter value deviates from the value, 1.039 cGyh(-1)U(-1), recommended in the TG-43 report. The calculated radial dose function for the Symmetra new seed is more penetrating than that of the model 6711 seed (by 20% at 5 cm distance) but agrees closely (within statistical errors) with that of the model 6702 seed up to distances of 10 cm. The anisotropy function for the seed is also close to that for the 6702 seed with a tendency of somewhat more pronounced anisotropy (lower values at small angles from the longitudinal axis). Compared to the model 6711 seed, the Symmetra new seed is more isotropic. The anisotropy constants (the anisotropy function averaged with respect to angle and distance) for the three seed models are within 2%.
机译:利用蒙特卡洛光子传输计算,对用于永久性前列腺植入物的新125I种子Symmetra 1251种子模型I25.S06进行了剂量学研究。 AAPM任务组43(TG-43)报告建议的所有剂量均已计算。确定的数量是剂量率常数,径向剂量函数,各向异性函数,各向异性因子和各向异性常数。考虑了最近(1999年1月)修订的NIST(美国国家标准技术研究院)1251标准,用于空气比释动能强度校准以及更新的相互作用横截面数据。为了比较,对竞争模型6702源进行了计算。 Symmetra和6702型种子的两个种子的计算剂量率常数分别为1.010和1.016 cGyh(-1)U(-1)。后者的值与TG-43报告中建议的值1.039 cGyh(-1)U(-1)不同。 Symmetra新种子的径向剂量函数比6711种子的径向剂量函数更具穿透性(在5 cm距离处增加20%),但在10 cm距离内与6702种子的径向剂量函数紧密相关(在统计误差范围内) 。种子的各向异性函数也接近于6702种子的各向异性函数,具有某种程度上更加各向异性的趋势(与纵轴成小角度的值较小)。与6711型种子相比,Symmetra新种子更具各向同性。三个种子模型的各向异性常数(相对于角度和距离平均的各向异性函数)在2%以内。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号