Comparison of real beat-to-beat signals with commercially available 4 Hz sampling on the evaluation of foetal heart rate variability

机译:实际心跳信号与市售4 Hz采样对胎儿心率变异性评估的比较

获取原文
获取原文并翻译 | 示例
           

摘要

Evaluation of foetal heart rate (FHR) variability is an essential part of foetal monitoring, but a precise quantification of this parameter depends on the quality of the signal. In this study, we compared real FHR beat-to-beat signals with 4 Hz sampling provided by commercial foetal monitors on linear and nonlinear indices and analysed their clinical implications. Simultaneous acquisition of beat-to-beat signals and their 4 Hz sampling rate counterparts was performed using a scalp electrode, during the last hour of labour in 21 fetuses born with an umbilical artery blood (UAB) pH ≥ 7.20 and 6 born with an UAB pH < 7.20. For each case, the first and last 10 min segments were analysed, using time and frequency domain linear, and nonlinear FHR indices, namely mean FHR, low frequency, high frequency, approximate, sample and multiscale entropy. Significant differences in variability indices were found between beat-to-beat and 4 Hz sampled signals, with a lesser effect seen with 2 Hz sampling. These differences did not affect physiological changes observed during labour progression, such as decreased entropy and linear time domain indices, and increased frequency domain indices. However, significant differences were found in the discrimination between fetuses born with different UAB pHs, with beat-to-beat sampling providing better results in linear indices and 4 Hz sampling better results in entropy indices. In conclusion, different FHR sampling frequencies can significantly affect the quantification of variability indices. This needs to be taken into account in the interpretation of FHR variability and in the development of new equipment.
机译:胎儿心率(FHR)变异性的评估是胎儿监测的重要组成部分,但是此参数的精确量化取决于信号的质量。在这项研究中,我们将真实的FHR逐搏信号与商用胎儿监护仪提供的4 Hz采样的线性和非线性指标进行了比较,并分析了它们的临床意义。在头一个工作的最后一个小时中,用头皮电极同时采集心跳信号及其4 Hz采样率对应信号,对21名胎儿脐带血(UAB)pH≥7.20的胎儿和6名UAB胎儿进行分娩pH值<7.20。对于每种情况,使用时域和频域线性和非线性FHR指数(即平均FHR,低频,高频,近似,样本和多尺度熵)对前10分钟和后10分钟进行分析。在逐拍和4 Hz采样信号之间发现了变异性指数的显着差异,而在2 Hz采样中观察到的影响较小。这些差异不会影响分娩过程中观察到的生理变化,例如熵和线性时域指数的降低以及频域指数的提高。但是,发现在具有不同UAB pH的胎儿之间的区分上存在显着差异,逐跳采样在线性指标中提供了更好的结果,而4 Hz采样在熵指标中提供了更好的结果。总之,不同的FHR采样频率可以显着影响变异性指标的量化。在解释胎心率变异性和开发新设备时,必须考虑到这一点。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号