首页> 外文期刊>Medical and Biological Engineering and Computing: Journal of the International Federation for Medical and Biological Engineering >Comparative analysis of power-line interference between two- or three-electrode biopotential amplifiers.
【24h】

Comparative analysis of power-line interference between two- or three-electrode biopotential amplifiers.

机译:两电极或三电极生物电势放大器之间的电源线干扰比较分析。

获取原文
获取原文并翻译 | 示例
           

摘要

Differential input amplifiers are commonly used in the measurement of biopotential signals. This is because of their ability to reject common-mode signals, for example power-line interference. Rejection is further enhanced by referencing the subject to the instrumentation system. One such referencing technique uses a third subject electrode connected either to circuit common or to ground. However, these techniques are rarely explained or quantified in the literature. Four recording configurations are examined, isolated and non-isolated amplifiers with two and with three electrodes, and existing models are extended. From a brief assessment of power-line interference sources, it is concluded that only displacement currents induced into the leads and the body need be considered. The effect of these on the four configurations is analysed using SPICE. Results show that interference from displacement currents induced into the leads and into the body is dependent upon the recording configuration used and is significantly lower in three-electrode than in two-electrode amplifiers. Isolation also reduces levels of interference. It is concluded that the choice of referencing may affect interference levels on the biopotential signal significantly. Experimental and further analytical work is planned to confirm this.
机译:差分输入放大器通常用于生物电势信号的测量。这是因为它们具有抑制共模信号(例如电力线干扰)的能力。通过将对象引向仪器系统,可以进一步提高拒绝感。一种这样的参考技术使用连接到公共电路或接地的第三对象电极。但是,这些技术很少在文献中得到解释或量化。检查了四种记录配置,具有两个和三个电极的隔离和非隔离放大器,并扩展了现有模型。通过对电力线干扰源的简要评估,可以得出结论,仅需要考虑感应到引线和主体中的位移电流。使用SPICE分析了这些对四种配置的影响。结果表明,感应到引线和人体中的位移电流的干扰取决于所使用的记录配置,并且在三电极放大器中比在三电极放大器中明显更低。隔离还可以降低干扰水平。结论是参考的选择可能会严重影响生物电势信号的干扰水平。计划进行实验和进一步的分析工作以证实这一点。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号