摘要

Since the early 2000s, a small but rapidly increasing number of patients with breast cancer have been treated with proton beams. Some of these patients have had breast prostheses or tissue expanders in place during their courses of treatment. Procedures must be implemented to plan the treatments of these patients. The density, kilovoltage x-ray computed tomography numbers (kVXCTNs), and proton relative linear stopping powers (pRLSPs) were calculated and measured for several test sample devices. The calculated and measured kVXCTNs of saline were 1% and 2.4% higher than the values for distilled water while the calculated RLSP for saline was within 0.2% of the value for distilled water. The measured kVXCTN and pRLSP of the silicone filling material for the test samples were approximately 1120 and 0.935, respectively. The conversion of kVXCTNs to pRLSPs by the treatment planning system standard tissue conversion function is adequate for saline-filled devices but for silicone-filled devices manual reassignment of the pRLSPs is required.
机译:自2000年代初以来,少数但迅速增加的乳腺癌患者接受了质子束治疗。这些患者中有一些在治疗过程中已经安装了乳房假体或组织扩张器。必须执行程序以计划这些患者的治疗。计算并测量了几个测试样品设备的密度,千伏X射线计算机断层扫描数(kVXCTN)和质子相对线性停止功率(pRLSPs)。计算和测量的盐水kVXCTN比蒸馏水的值分别高1%和2.4%,而计算得出的盐水RLSP在蒸馏水值的0.2%以内。对于测试样品,测得的有机硅填充材料的kVXCTN和pRLSP分别约为1120和0.935。通过治疗计划系统标准组织转换功能将kVXCTN转换为pRLSP,对于充满生理盐水的设备就足够了,但是对于充满有机硅的设备,则需要手动重新分配pRLSP。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号