Assessment of MR compatibility of a PET insert developed for simultaneous multiparametric PET/MR imaging on an animal system operating at 7 T.

机译:评估PET插入件的MR兼容性,该PET插入件用于在7 T的动物系统上同时进行多参数PET / MR成像。

获取原文
获取原文并翻译 | 示例
           

摘要

The combination of positron emission tomography and MR in one system is currently emerging and opens up new domains in the functional examinations of living systems. This article reports on relevant influences of a positron emission tomography insert on MR imaging. The basic conditions of main magnetic field and RF field homogeneity were measured as well as image quality and signal-to-noise ratio when applying the usual MR sequence types including echo-planar techniques. Moreover, the influence of the positron emission tomography insert on the RF noise level and on RF interferences was measured by comparing results achieved with and without the positron emission tomography insert. The temporal stability of EPI imaging with and without the positron emission tomography insert was assessed. Small but significant decreases in the signal-to-noise ratio were revealed when the positron emission tomography insert was present, whereas B(0) and B(1) homogeneity as well as RF noise level were not adversely affected. A higher signal intensity drift was found for EPI imaging studies; however, this can be compensated by post processing. In summary, this study shows that positron emission tomography inserts can be designed for and used within an MR system practically, without substantially affecting the MR image quality.
机译:正将一种正电子发射断层扫描和MR结合在一个系统中正在兴起,并在生命系统的功能检查中开辟了新领域。本文报道了正电子发射断层扫描插入物对MR成像的相关影响。当应用包括回波平面技术在内的常规MR序列类型时,测量了主磁场和RF场均匀性的基本条件,以及图像质量和信噪比。此外,通过比较使用和不使用正电子发射断层摄影插入物的结果,可以测量正电子发射断层摄影插入物对RF噪声水平和RF干扰的影响。评估有和没有正电子发射断层扫描插入物的EPI成像的时间稳定性。当存在正电子发射断层扫描插入物时,发现信噪比小幅但显着降低,而B(0)和B(1)的均匀性以及RF噪声水平并未受到不利影响。对于EPI成像研究,发现较高的信号强度漂移;但是,这可以通过后期处理来补偿。总而言之,这项研究表明,实际上可以为MR系统设计并在MR系统中使用正电子发射断层扫描插件,而不会实质性地影响MR图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号