Multicoil Dixon chemical species separation with an iterative least-squares estimation method.

机译:采用迭代最小二乘估计方法的Multicoil Dixon化学物种分离。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

This work describes a new approach to multipoint Dixon fat-water separation that is amenable to pulse sequences that require short echo time (TE) increments, such as steady-state free precession (SSFP) and fast spin-echo (FSE) imaging. Using an iterative linear least-squares method that decomposes water and fat images from source images acquired at short TE increments, images with a high signal-to-noise ratio (SNR) and uniform separation of water and fat are obtained. This algorithm extends to multicoil reconstruction with minimal additional complexity. Examples of single- and multicoil fat-water decompositions are shown from source images acquired at both 1.5T and 3.0T. Examples in the knee, ankle, pelvis, abdomen, and heart are shown, using FSE, SSFP, and spoiled gradient-echo (SPGR) pulse sequences. The algorithm was applied to systems with multiple chemical species, and an example of water-fat-silicone separation is shown. An analysis of the noise performance of this method is described, and methods to improve noise performance through multicoil acquisition and field map smoothing are discussed.
机译:这项工作描述了一种多点狄克逊脂肪-水分离的新方法,该方法适用于需要短回波时间(TE)增量的脉冲序列,例如稳态自由进动(SSFP)和快速自旋回波(FSE)成像。使用迭代线性最小二乘法将水和脂肪图像从以短TE增量获取的源图像分解,可以获得具有高信噪比(SNR)和水和脂肪均匀分离的图像。该算法以最小的附加复杂性扩展到多线圈重构。从在1.5T和3.0T处采集的源图像显示了单线圈和多线圈脂肪水分解的示例。使用FSE,SSFP和损坏的梯度回波(SPGR)脉冲序列显示了膝盖,脚踝,骨盆,腹部和心脏的示例。该算法已应用于具有多种化学物种的系统,并显示了水-脂肪-硅树脂分离的一个示例。对该方法的噪声性能进行了分析,并讨论了通过多线圈采集和场图平滑来改善噪声性能的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号