首页> 外文期刊>Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine >A geometrically adjustable 16-channel transmit/receive transmission line array for improved RF efficiency and parallel imaging performance at 7 Tesla.
【24h】

A geometrically adjustable 16-channel transmit/receive transmission line array for improved RF efficiency and parallel imaging performance at 7 Tesla.

机译:几何可调的16通道发射/接收传输线阵列,可提高RF效率和7 Tesla的并行成像性能。

获取原文
获取原文并翻译 | 示例
           

摘要

A novel geometrically adjustable transceiver array system is presented. A key feature of the geometrically adjustable array was the introduction of decoupling capacitors that allow for automatic change in capacitance dependent on neighboring resonant element distance. The 16-element head array version of such an adjustable coil based on transmission line technology was compared to fixed geometry transmission line arrays (TLAs) of various sizes at 7T. The focus of this comparison was on parallel imaging performance, RF transmit efficiency, and signal-to-noise ratio (SNR). Significant gains in parallel imaging performance and SNR were observed for the new coil and attributed to its adjustability and to the design of the individual elements with a three-sided ground plane.
机译:提出了一种新颖的几何可调收发器阵列系统。几何可调阵列的关键特征是引入了去耦电容器,该电容器可根据相邻谐振元件的距离自动改变电容。将这种基于传输线技术的可调线圈的16元头阵列版本与7T时各种尺寸的固定几何传输线阵列(TLA)进行了比较。该比较的重点是并行成像性能,RF传输效率和信噪比(SNR)。对于新线圈,观察到并行成像性能和SNR的显着提高,这归因于其可调整性以及具有三侧接地层的单个元件的设计。