首页> 外文期刊>Canadian journal of anesthesia: Journal canadien d'anesthesie >PVC tracheal tubes exert forces and pressures seven to ten times higher than silicone or armoured tracheal tubes - an in vitro study: (Les tubes tracheaux de PVC exercent des forces et des pressions de sept a dix fois plus elevees que les tubes en si
【24h】

PVC tracheal tubes exert forces and pressures seven to ten times higher than silicone or armoured tracheal tubes - an in vitro study: (Les tubes tracheaux de PVC exercent des forces et des pressions de sept a dix fois plus elevees que les tubes en si

机译:PVC气管导管施加的压力和压力是硅胶或铠装气管导管的七到十倍-一项体外研究:(PVC气管导管施加的压力和压力是硅胶或铠装气管导管的七到十倍

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: Many types of tracheal tubes (TT) including silicone, polyvinylchloride (PVC) and armoured have been used for blind tracheal intubation (TI) via the intubating laryngeal mask airway (ILMA) and may cause trauma to the airway. We examined the maximal in vitro forces and pressures exerted by the tip of various TT as it exits the ILMA. METHODS: Silicone, PVC and armoured TT were studied. A #5 ILMA was secured on a wooden platform. With the use of a Harvard pump, force was applied to push the TT through the ILMA at 0.34 cm*sec(-1). Forces exerted to push the TT and forces exerted by the TT tips on distal objects were calculated using proximal and distal pressure manometres. The areas of contact between the distal TT tips and the distal objects were measured by planimetry of an imprint. The final pressures exerted by the TT tips on a fixed distal object were calculated by dividing the forces exerted by the areas of contact. RESULTS: When compared to silicone and armoured TT, PVC TT exerted seven to ten times higher forces and pressures on distal objects. (P < 0.05). Heating PVC TT and inserting PVC TT with reverse curvature to the ILMA did not decrease the forces and pressures exerted by the distal tip. CONCLUSION: The high forces and pressures exerted by PVC TT may theoretically contribute to increased morbidity to patients' airway and esophagus. Caution should be exercised before attempting blind TI via the ILMA with a PVC TT.
机译:目的:许多类型的气管导管(TT),包括硅树脂,聚氯乙烯(PVC)和铠装,已通过插管式喉罩气道(ILMA)用于气管插管(TI),可能会对气道造成伤害。我们检查了各种TT尖端离开ILMA时所施加的最大体外作用力和压力。方法:研究了有机硅,聚氯乙烯和铠装TT。 #5 ILMA固定在木制平台上。使用哈佛泵,施加力以0.34 cm * sec(-1)的力将TT推过ILMA。使用近端压力计和远端压力计来计算推动TT的力和TT尖端对远端对象的力。远端TT尖端和远端对象之间的接触面积通过压印的平面法进行测量。通过将TT尖端施加在固定的远端物体上的最终压力通过将施加的力除以接触面积来计算。结果:与硅胶和铠装TT相比,PVC TT对远端物体施加的力和压力高出七到十倍。 (P <0.05)。加热PVC TT并将曲率相反的PVC TT插入ILMA并不会降低远端尖端施加的力和压力。结论:PVC TT施加的巨大压力和压力理论上可能会增加患者气道和食道的发病率。在通过ILMA与PVC TT进行盲TI尝试之前,应谨慎行事。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号