The scavenging of volatile anesthetic agents in the cardiovascular intensive care unit environment: A technical report

机译:在心血管重症监护病房环境中清除挥发性麻醉剂的技术报告

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Purpose: The use of volatile-based sedation within critical care environments has been limited by difficulties of drug administration and safety concerns over environment pollution and staff exposure in an intensive care unit (ICU) with no scavenging. The aim of this study was to develop a simple scavenging system to be used with the Anesthesia Conserving Device (AnaConDa?) and to determine whether or not ambient concentrations of residual anesthetic are within current acceptable limits. Technical features: The scavenging system consists of two Deltasorb ? canisters attached to the ICU ventilator in series. AnaConDa is a miniature vaporizer designed to provide volatile-based sedation within an ICU. The first ten patients recruited into a larger randomized trial assessing outcomes after elective coronary graft bypass surgery were sedated within the cardiac ICU using either isoflurane or sevoflurane. Sedation was guided by the Sedation Agitation Scale, resulting in an end-tidal minimum anesthetic concentration of volatile agent ranging from 0.1-0.3. At one hour post ICU admission, infrared photometric analysis was used to assess environmental contamination at four points along the ventilator circuit and scavenging system and around the patient's head. All measurements taken within the patient's room were below 1 part per million, which satisfies criteria for occupational exposure. Conclusions: This study shows that volatile agents can be administered safely within critical care settings using a simple scavenging system. Our scavenging system used in conjunction with the AnaConDa device reduced the concentration of environmental contamination to a level that is acceptable to Canadian standards and standards in most Western countries and thus conforms to international safety standards. The related clinical trial was registered at www.clinicaltrials.gov (NCT01151254).
机译:目的:由于药物管理的困难以及对环境污染的安全担忧以及在没有清除措施的重症监护病房(ICU)中的工作人员接触,限制了在重症监护环境中使用基于挥发性的镇静剂。这项研究的目的是开发一种与麻醉保存装置(AnaConDa?)一起使用的简单清除系统,并确定残留麻醉剂的环境浓度是否在当前可接受的范围内。技术特点:清除系统由两个Deltasorb?串联连接到ICU呼吸机的碳罐。 AnaConDa是一款微型蒸发器,旨在在ICU中提供基于挥发性的镇静剂。在参加选择性大面积冠状动脉搭桥手术后评估结局的较大随机试验中,前十名患者使用异氟烷或七氟醚在心脏ICU中镇静。在镇静搅拌量表的指导下进行镇静,导致潮气末的挥发性药物最低麻醉浓度为0.1-0.3。 ICU入院后一小时,使用红外光度分析法评估呼吸机回路和扫气系统以及患者头部周围四个点的环境污染。在患者房间内进行的所有测量均低于百万分之一,这符合职业暴露标准。结论:这项研究表明,可以使用简单的清除系统在重症监护环境中安全地施用挥发性药物。我们的清除系统与AnaConDa设备结合使用,将环境污染的浓度降低到加拿大标准和大多数西方国家的标准可以接受的水平,从而符合国际安全标准。相关的临床试验已在www.clinicaltrials.gov(NCT01151254)上注册。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号