A method for the calibration of closely spaced multiple hot-film sensor (MHFS) arrays of nickel-on-polyimide-substrate type was presented in this paper. The hot-film sensors were calibrated individually and in situ against the values of the wall-shear stresses determined from boundary-layer velocity profiles (developed on the front portion of a circular cylinder) through a specially designed wall-mounted hot-wire probe. The accuracy and feasibility of the present method was examined through the subsequent measurements of the skin-friction distributions on a circular cylinder (including the removal and bonding of the calibrated MHFS from a 10-cm diameter calibration cylinder and onto the surface of a 8-cm diameter test cylinder) and comparison of their consistency with the published results. The present MHFS calibration method is important in the non-intrusive multipoint measurements of the mean skin friction along the surfaces of two-dimensional aerodynamic objects of interest. [References: 24]
机译:本文提出了一种校准聚酰亚胺镍基衬底上的近距离多重热膜传感器(MHFS)阵列的方法。通过专门设计的壁挂式热线探头,根据从边界层速度分布图(在圆柱体的前部形成)确定的壁切应力值,分别对热膜传感器进行了原位校准。通过随后在圆柱上的皮肤摩擦分布测量(包括从直径为10 cm的校准圆筒上将校准的MHFS去除并粘结到8毫米表面上的MHFS上),检验了本方法的准确性和可行性。厘米直径的测试缸),并将其与已发布结果的一致性进行比较。当前的MHFS校准方法在沿着感兴趣的二维空气动力学物体表面的平均皮肤摩擦的非侵入式多点测量中很重要。 [参考:24]