Inter-scan reproducibility of coronary calcium measurement using Multi Detector-Row Computed Tomography (MDCT).

机译:使用多探测器行计算机断层扫描(MDCT)进行冠状动脉钙测量的扫描间可重复性。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: To assess inter-scan reproducibility of coronary calcium measurements obtained from Multi Detector-Row CT (MDCT) images and to evaluate whether this reproducibility is affected by different measurement protocols, slice thickness, cardiovascular risk factors and/or technical variables. DESIGN: Cross-sectional study with repeated measurements. MATERIALS AND METHODS: The study population comprised 76 healthy women. Coronary calcium was assessed in these women twice in one session using 16-MDCT (Philips Mx 8000 IDT 16). Images were reconstructed with 1.5 mm slice thickness and 3.0 mm slice thickness. The 76 repeated scans were scored. The Agatston score, a volume measurement and a mass measurement were assessed. Reproducibility was determined by estimation of mean, absolute, relative difference, the weighted kappa value for agreement and the Intra-class correlation coefficient (ICCC). RESULTS: Fifty-five participants (72.4%) had a coronary calcification of more than zero in Agatston (1.5 mm slice thickness). The reproducibility of coronary calcium measurements between scans in terms of ranking was excellent with Intra-class correlation coefficients of >0.98, and kappa values above 0.80. The absolute difference in calcium score between scans increased with increasing calcium levels, indicating that measurement error increases with increasing calcium levels. However, no relation was found between the mean difference in scores and calcium levels, indicating that the increase in measurement error is likely to result in random misclassification in calcium score. Reproducibility results were similar for 1.5 mm slices and for 3.0 mm slices, and equal for Agatston, volume and mass measurements. CONCLUSION: Inter-scan reproducibilility of measurement of coronary calcium using images from MDCT is excellent, irrespective of slice thickness and type of calcium parameter.
机译:目的:评估从多排螺旋CT(MDCT)图像获得的冠状动脉钙测量值的扫描间可重复性,并评估这种可重复性是否受到不同的测量方案,切片厚度,心血管危险因素和/或技术变量的影响。设计:横断面研究,重复测量。材料与方法:研究人群包括76名健康女性。使用16-MDCT(Philips Mx 8000 IDT 16)在一个疗程中两次评估这些女性的冠状动脉钙。用1.5毫米切片厚度和3.0毫米切片厚度重建图像。对76次重复扫描进行了评分。评估了Agatston评分,体积测量和质量测量。通过估计均值,绝对值,相对差,协议的加权kappa值和类内相关系数(ICCC)来确定可重复性。结果:55名参与者(72.4%)的阿格斯顿(1.5毫米厚)的冠状动脉钙化大于零。两次扫描之间冠状动脉钙测量值的重现性极好,类内相关系数> 0.98,kappa值大于0.80。扫描之间钙分数的绝对差异随钙水平的升高而增加,表明测量误差随钙水平的升高而增加。但是,在评分的平均差异与钙水平之间未发现任何关系,这表明测量误差的增加很可能导致钙评分的随机错误分类。 1.5 mm切片和3.0 mm切片的重现性结果相似,而Agatston,体积和质量测量结果均相同。结论:无论切片厚度和钙参数类型如何,使用MDCT图像测量冠状动脉钙的扫描间可重复性均非常好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号